NeuLog NUL-223 Manual Download Page 1

NEULOG DROP COUNTER LOGGER SENSOR GUIDE

 

 

 

NeuLog drop counter logger sensor 
NUL-223 

 
The  NeuLog  drop  counter  sensor  can  be  used  for  any  science  experiment  which 
utilizes fluid drip rates, particularly in the fields of Chemistry and Biology. 
 
The sensor comes pre-calibrated so you can start experimentation right out of the box 
using this guide.  
 
The NeuLog drop counter sensor excels in the laboratory but can also be used in the 
field  for  a  number  of  experiments  in  various  configurations.  For  example,  titrations 
become  much  more  accurate  as  you  can  determine  both  the  total  number  of  drops 
and the total volume which has passed through the drop counter sensor.   
 
The drop counter sensor's measurement units are: 

 

Drops:  The  number  of  times  a  drop  of  liquid  passes  through  the  drop 
counter sensor. 

 

Milliliters  (mL):  The  SI  unit  of  volume.  The  NeuLog  drop  counter  sensor 
calculates the rough volume of drops which pass through it. 

 

Setting up the drop counter sensor: 

 
There  are  a  number  of  ways  to  set  up  the  NeuLog  drop  counter  sensor  to  give 
accurate  readings.    The  drop  counter  gate  includes  three  threaded  holes  and  a 
threaded  utility  rod  which  can  be  attached  to  various  lab  equipment  to  offer  many 
different orientations. 
The most important aspect while setting up the drop counter sensor is to ensure that 
the  drops  of  liquid  are  passing  through  the  sensor  correctly.  As  in  the  figure  below, 
you  will  want  to  orient  the  sensor  so  that  drops  pass  through  the  infrared  beam 
transmitted between the ends of the drop counter sensor. 
 

 

Quick start procedure: 

 

PC or Mac Computer 

 
Materials needed: 

 

NUL-223 Drop Counter Sensor 

 

USB-200 USB Module 

 

USB to mini USB cable (included with the USB-200 module) 

 
Your drop counter sensor needs to be connected to a USB-200 module. The USB-200 
module  then connects  to  a  computer  via a  USB to mini-USB  cable.  Please  note  that 
you cannot plug the drop counter sensor directly into the computer. 
 
The  browser  based  application  can  be  downloaded  for  free  at 

www.NeuLog.com

  as 

well as a full application user guide. 
Choose "Downloads" on the main menu and then choose "Software and Application". 

 
 
 
Procedure: 

 
1. 

Install the NeuLog application. 

2. 

Connect the USB-200 module to the PC or Mac. 

3. 

Connect  the  drop  counter  sensor  to  the  USB-200  module  (they  directly  plug 
together).  Please note that no calibration is required for this sensor. 

4. 

Open the NeuLog application by clicking on the shortcut on the screen. 

5. 

Once a drop counter sensor module box appears on the left side of the screen 
the probe has been automatically identified and you can begin experimentation. 

6. 

If the drop counter sensor is not automatically identified then click the “Search 
for sensors” icon to find the sensor. 

7. 

Select th

e “On-line experiment” icon; this will open a graph below. 

8. 

Click  on  the  “Module  setup”  button  located  in  the  drop  counter  sensor  module 
box to change the sensor settings if need be. 

9. 

Click on the "Experiment setup" button to change the experiment settings if need 
be (experiment duration for example). 

10.  The drop counter sensor will give a live reading in the module box to the left of 

the screen while plugged in. 

11. 

To run an experiment and collect data click the “Run experiment” icon. 

12.  To end data collection early, 

click the “Stop experiment” icon. 

 

 

Summary of Contents for NUL-223

Page 1: ... Quick start procedure PC or Mac Computer Materials needed NUL 223 Drop Counter Sensor USB 200 USB Module USB to mini USB cable included with the USB 200 module Your drop counter sensor needs to be connected to a USB 200 module The USB 200 module then connects to a computer via a USB to mini USB cable Please note that you cannot plug the drop counter sensor directly into the computer The browser b...

Page 2: ...nto the URL bar then wait for 30 60 seconds 10 You will see a Control mode icon in the application click on it 11 The application will then load a new screen and begin to auto detect the sensor s this can take a minute You can stop the search when the sensor is found 12 If the application does not auto detect the sensor s select Search for sensors 13 Once the sensor is found you will see a module ...

Page 3: ... Procedure 1 Connect the USB 200 to a PC Mac tablet or smart device 2 Connect the drop counter sensor to a USB 200 module they directly plug together 3 Open the NeuLog application 4 Click the Tools icon 5 Click Restore sensor s factory defaults on the menu Included with the sensor NeuLog NUL 223 Drop Counter Sensor Guide this document Durable plastic drop counter probe attached directly to the sen...

Page 4: ...quid s opacity or color the infrared light bends away from the detector Each time there is a break in the infrared beam a drop is counted Maintenance and storage Never submerge the NeuLog plastic body in any liquid Do not allow liquid into the drop counter sensor s body and probe After use gently wipe away any foreign material from the drop counter sensor Store in a box at room temperature out of ...

Reviews: