NeuLog NUL-210 Manual Download Page 3

NEULOG PRESSURE SENSOR GUIDE

 

 

 

14.

 

Click on the “Module setup” button located on the temperature sensor 
icon in the module window to change the temperature probe settings if 
need be. 

15.

 

Click  on  the  experiment  set  up  button  to  change  the  experiment 
settings if need be (Experiment duration for example). 

16.

 

To  run  an  experiment  and  collect  data,  select  “On-Line”  experiment, 
then select "Run". 

17.

 

If  you  have  a  newer  generation  WIFI-201  (with  the  USB  icon  on  the 
cover)  this  can  be  used  as  a  USB-200,  meaning  you  can  tether  the 
probes  directly  to  the  computer  using  the  WIFI-201.  To  engage 
operation in this fashion, press 3 times on the panel key. Repeating this 
operation will return the unit to WiFi mode. 

 

Operation with NeuLog viewer 

Materials needed: 

 

NUL-210 Pressure Sensor 

 

VIEW-101 Viewing Module 

 

BAT-200 Battery 

 

Procedure: 

 
1.

 

Connect the pressure sensor to the left side of the viewer. 

2.

 

Connect the BAT-200 module to the right side of the viewer. 

3.

 

The  VIEW-101  will  auto  detect  the  sensor  and  you  will  see  it  appear 
on the left side of the screen. 

4.

 

Once the sensor appears it will be monitoring data in real time. 

5.

 

To run an experiment and collect data click the run button (little green 
person). 

 
 
 

 

Off-line experiments 

 
(Off-line  experiments  are  for  when  you  do  not  have  a  sensor  connected 
directly to a computer, tablet, smartphone, or NeuLog viewer). 
 

Materials needed: 
 

 

NUL-210 Pressure Sensor 

 

BAT-200 Battery 

 

Materials needed to configure your offline experiment: 
 

 

USB-200 Module or WIFI-201 module or VIEW-101 Module 

 

USB to mini USB cable (included with the USB-200 module). 

 

Procedure: 

 
1.

 

Connect the pressure sensor directly to the left side of a charged BAT-
200 module. 

2.

 

When  ready  to  collect  data,  press  the  “Start/Stop”  button  on  the 
pressure  sensor;  a  red  light  will  turn  on  for  the  duration  of  the 
experiment. 

3.

 

To  change  the  experiment's  settings,  first  connect  to  your  chosen 
device  (PC,  Mac,  tablet,  smart  device)  and  select  the  “Module  setup” 
button.  (For  more  information  on  how  to  change  experiment  settings 
view  the  quick  start  procedure  section  for  your  chosen  device  in  this 
document.) 

4.

 

After  the  experiment  has  concluded  (when  the  “Start/Stop”  button  is 
pressed again or the data collection period ends) connect the sensor to 
your  chosen  device  (you  can  find  full  instructions  in  each  device’s 
quick start procedure section). 

5.

 

Open the NeuLog software. 

Reviews: