
NEULOG PRESSURE SENSOR GUIDE
14.
Click on the “Module setup” button located on the temperature sensor
icon in the module window to change the temperature probe settings if
need be.
15.
Click on the experiment set up button to change the experiment
settings if need be (Experiment duration for example).
16.
To run an experiment and collect data, select “On-Line” experiment,
then select "Run".
17.
If you have a newer generation WIFI-201 (with the USB icon on the
cover) this can be used as a USB-200, meaning you can tether the
probes directly to the computer using the WIFI-201. To engage
operation in this fashion, press 3 times on the panel key. Repeating this
operation will return the unit to WiFi mode.
Operation with NeuLog viewer
Materials needed:
NUL-210 Pressure Sensor
VIEW-101 Viewing Module
BAT-200 Battery
Procedure:
1.
Connect the pressure sensor to the left side of the viewer.
2.
Connect the BAT-200 module to the right side of the viewer.
3.
The VIEW-101 will auto detect the sensor and you will see it appear
on the left side of the screen.
4.
Once the sensor appears it will be monitoring data in real time.
5.
To run an experiment and collect data click the run button (little green
person).
Off-line experiments
(Off-line experiments are for when you do not have a sensor connected
directly to a computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-210 Pressure Sensor
BAT-200 Battery
Materials needed to configure your offline experiment:
USB-200 Module or WIFI-201 module or VIEW-101 Module
USB to mini USB cable (included with the USB-200 module).
Procedure:
1.
Connect the pressure sensor directly to the left side of a charged BAT-
200 module.
2.
When ready to collect data, press the “Start/Stop” button on the
pressure sensor; a red light will turn on for the duration of the
experiment.
3.
To change the experiment's settings, first connect to your chosen
device (PC, Mac, tablet, smart device) and select the “Module setup”
button. (For more information on how to change experiment settings
view the quick start procedure section for your chosen device in this
document.)
4.
After the experiment has concluded (when the “Start/Stop” button is
pressed again or the data collection period ends) connect the sensor to
your chosen device (you can find full instructions in each device’s
quick start procedure section).
5.
Open the NeuLog software.