Legend
1. Measuring clamp
2. Carbide measuring surfaces
3. Measuring spindle
4. Locking device
5. Scale
6. Ratchet
7. Heat protection plate
8. ON/OFF - SET button
9. LCD display
10. Battery compartment (rear side)
Specification
Measuring range:
0…25 / 25…50 / 50…75 / 75…100 mm
Resolution:
0.001 mm
Measuring force:
5…10 N
Operating temperature:
0…40°C
Power:
Battery SR44 / 1.55V
Protection class:
IP65
Properties
ON/OFF:
Switch display on/off
SET:
Set pre-selection value (zeros)
Auto-Power OFF:
Automatic shutdown
Operation
Switch on/off:
Short press of ON/OFF button
Set pre-selection value (zeros): Long press of SET button
DIGITAL MICROMETER
Operating instructions
Safety instructions
• Clean measuring surfaces before use
• Use the micrometer only at the permissible operating temperature
• Protect the micrometer from mechanical damage
• The electronics must not be exposed to a strong magnetic field
• The electronics must not be opened
• If the battery icon flashes in the display - change the battery
• Dispose of battery and micrometer properly
Instructions for care
A soft cloth can be used for cleaning the surfaces.
Do not use petrol, acetone or isopropanol for cleaning under any circumstances.
Remove the battery if the instrument is not going to be used for an extended period.
MAN_BR
W
_div
. March_20_art. BR
W
1151
49
Repair
If the instrument is defective, please contact our
service centre → [email protected]