Questions or issues?
Let us help!
Call:
(201) 343-5202, Ext. 1
Installation Guide
Part #:
70-0285
(See DSU-H Instructions for Use)
Questions or issues?
Let us help!
Call:
(201) 343-5202, Ext. 1
OFF
OFF
DISCONNECT
DISCONNECT
1
2
3
READ the entire installation guide and
precautions before starting.
WASH hands before starting installation.
WEAR gloves before handling filter (do not
remove blister pack).
Turn OFF the water supply valves and OPEN
faucet.
DISCONNECT, REMOVE, and DISCARD the
current flex lines on both sides of sink.
REMOVE and DISCARD the aerator (connected
to spout).
CONNECT
CONNECT
SINK
CONNECT
INLET
BLEACH
OPEN
INSERT
4
5
6
CONNECT the faucet stem flex line to the faucet
stem by the 1/2” NPSF fitting.
CONNECT the quick connector with bushing
onto the other end of the faucet stem flex line.
Perform steps for cold and hot sides of sink.
Clorox
®
Healthcare Bleach Germicidal Cleaner is
the recommended disinfectant choice.
CONNECT the outlet of the faucet disinfection
unit to the faucet stem flex line quick connector.
PLACE the inlet hose of unit into disinfectant
container.
Ensure that faucet valve is open.
INSERT a piece of gauze at end of spout.
DSU-H Sink
AVOID contact between the filter and the cleaning or sanitizing agents.
(1)
DO NOT remove blister pack from filter until indicated to protect from
contamination.
SINK
SINK
OFF
OFF
DISCONNECT
DISCONNECT
CONNECT
CONNECT
SINK
CONNECT
INLET
DISINFECTANT
SINK
OPEN
INSERT
(1) Medical grade plastics are more susceptible to cracking due to exposure to organic solvents. Alkyl
dimethyl benzyl ammonium chloride (ADBAC), glycols, alcohols, ketones, silicones, and acetones
are some examples of the leading organic solvents found in cleaning and sanitizing agents. Use of
cleaning/sanitizing agents on the filter housings is not required and should be avoided.