AccuPoint Advanced HC User’s Guide
12
SYNCING TEST RESULTS TO THE HC DATA MANAGER SOFTWARE
The AccuPoint Advanced HC reader can store up to 4,000 data records. The reader will
display the number of stored data records on the User ID screen at power-up. Once the
limit has been reached, the oldest record will be discarded as each new record is added.
To sync data to a computer:
1. Open the Data Manager program on the computer (if necessary, see the software
manual for program installation instructions).
2. Turn on the reader and allow it to move past the boot up screen. Select a
User
.
3. Connect the USB from the computer to the reader.
4. Any data that is on the instrument will automatically be presented for review.
5. When syncing is complete, a message is displayed indicating that data has been
successfully loaded into the HC Data Manager program. The program will ask if
the data in the reader should be deleted.
6. To erase the data from the reader, click Yes in the message box. Once data has
been erased, it cannot be retrieved. Click No to retain the data in the reader.
NOTE:
to avoid seeing the
Delete All Records
message, select
Auto Delete
After Upload
from the Settings tab.
7. Turn off the reader.
SETTING THRESHOLDS IN THE READER
The default thresholds come pre-programmed at 0–149 as pass, 150–299 as marginal
and above 300 as fail. Changing them can only be accomplished by uploading through the
Data Manager software. When a Test Plan is established, each site can be programmed to
its unique thresholds. Once the Test Plan is complete, uploading it to the reader through the
Download Thresholds command will establish the correct thresholds for each site.
If a Test Plan will not be used at a facility, and to simply operate the system in a pass-
fail mode, it will still be necessary to upload a simple test plan to not use the default
thresholds. Simply establish a single test site in the Test Plan Setup routine with its
appropriate thresholds and upload it to the reader through the Download Thresholds
command. This will change the threshold defaults in the reader to the values that have
been established.