3.8.2 Pin Description
3.8.3 External Control Connector - Technical description
1. Connector type : AMP, Part No: 5102617-5 14 contacts, double row.
2. Receptacle : AMP, Part No: 87631-9, Polarized, 14 contacts, double row.
3. Receptacle contacts : AMP, Part No: 87523-5 (strip form) or 87523-6 (loose piece).
4. Wires : AWG: 24 to 20.
5. Hand tool for wire connection : AMP, Part No: 90202-2.
6. Applicator for stripper/crimper machine : AMP, Part No: 466905-1
#
Name
1 RCVP
2 RCCP
3 VRFI
4 VCCP
5 VRFV
6 VCVP
7 OUTPUT
GOOD
8 P
9 ON/OFF
10 COM
11 + LS
12 + S
13 - LS
14 - S
Description
Resistive Constant Voltage programming
Resistive Constant Current programming
Reference voltage for Constant Current control. (connected to
VCCP pin except external voltage or resistive programming).
Input of the Constant Current control circuit.
Reference voltage for Constant Voltage control. (connected to
VCVP pin except external voltage or resistive programming).
Input of the Constant Voltage control circuit.
Open collector output. 0~0.8V during normal operation, open at
fault condition (OVP, FOLD and Over-Temp.).
Parallel connection pin.
On/Off control pin.
Common pin for programming, On/off control and Output Good
signal. Internally connected to - Sense.
Positive Local sensing pin.
Positive Sensing pin for remote sense operation.
Negative Local sensing pin.
Negative Sensing pin for remote sense operation.
4.4.10
4.4.10
4.4.10, 11, 12
4.4.10, 11, 12
4.4.10, 11, 12
4.4.10, 11, 12
4.4.15
4.4.13
4.4.8
4.4.10, 11, 12
4.4.8
3.7.3, 4, 5, 6
3.7.3, 4, 5, 6
3.7.3, 4, 5, 6
3.7.3, 4, 5, 6
Par.
Table 3-3: External Control connector pin description
CAUTION
COM pin (10) is connected internally to the - Sense potential. Connection of the COM to another
potential or other than shown in this manual may cause damage to the power supply.
3.9 REPACKAGING FOR SHIPMENT
To ensure safe transportation of the instrument, it is recommended to use the original package. The
original packaging material is reusable. If the original package is not available, contact the Nemic-
Lambda Sales or service facility near you for details on obtaining suitable packaging and shipping
information. Please attach a tag to the power supply describing the problem and specifying the owner,
model number and serial number of the power supply.
3.8.4 External Control Connector -
Default configuration connections
The external control connector is configured for Local sensing and no external On/Off control. Refer to
Fig: 3-6 below for the default configuration description. The default configuration is: Local sensing,
Rear panel On/Off disabled (On/Off can be controlled by the Front panel or by the serial communication
port), and output voltage and current are controlled by the Front panel or by the serial communication
port. For other configurations, remove the contact that should be modified, and replace it with a new
one with a suitable connection. Replacement contacts are supplied in the packaging of the power
supply.
Fig. 3-6:
External Control connector - default configuration
(ZUP rear panel view)
1
14
13
2
- LS
- S
+LS
+S
On/Off
COM
Output Good
P
VRFV
VCVP
VRFI
VCCP
RCVP
RCCP