Service Menu and Factory Default Settings
N5500 Service Manual 31
Default Settings and
Troubleshooting
11. Turn the monitor off, then power the monitor on again.
Note: The monitor must be powered off upon selecting “Done” to save any changes
into the monitor, and then the changes made to the Power on defaults will be in
effect next time the unit is powered up.
Note: If you do not select “YES” on save current values or back to factory defaults as
power-on default, the monitor will not save any changes and it will be
automatically powered off upon selecting “Done” to exit Service menu screen.
Table 9. Service menu
Menu Item
Response
Save Current Values as Power-On Default
YES, NO
Save Factory Default Setting as Power-On Default YES, NO
Alarm Suspend
On, Off
Alarm Silence Period
30, 60, 90, 120, 180 sec
Audible Alarm Type
GN924, IEC60601-1-8
AC Line Frequency
50Hz, 60Hz
Language
Chinese, Danish, Dutch,
English, Finnish, French,
German, Greek, Italian,
Norwegian, Polish, Portuguese,
Russian, Spanish, Swedish
NIBP Test Mode
Activate NIBP Test Mode screen
(see Table 11)
System Information
Activate System Information
screen (see Table 12)
Done
The monitor will be powered off
upon selecting “Done”, then any
changes will be in effect next
time the unit is powered up.
Save Current Values as Power-On Default
The current N5500 settings become the power-up defaults next time the unit is powered
up.
Save Factory Default Setting as Power-On Default
The N5500 factory default settings become the power-up defaults next time the unit is
powered up. See section 4.4 Factory Default Settings and Table 12.
Alarm Suspend
If “On” is chosen, the monitor may be able to invoke the Alarm Suspend mode by
pressing and holding the
Alarm silence/suspend switch
for 2 seconds. Some institutions
may wish to prevent Alarm Suspend from being invoked. If so, “Off” should be selected.
In this case, Alarm Suspend via the corresponding
numeric parameter menus
or the
alarm/limits menu
is also prevented from being invoked.