Audible Alarm Settings and Service Functions
38
If an invalid tone is heard instead of the triple beep, the current settings were
not changed. An invalid tone is a single low-pitched tone.
Note
:
Current values will not be stored in memory as institutional defaults if
power is interrupted before exiting this menu option.
Note
:
When the operator changes an alarm limit to a value other than a
institutional default value, a decimal point is displayed to the right of
the parameter whose alarm limit was changed.
Menu Item 10 (Restore Factory Default Settings as Power-On Default)
Menu item 10 resets the N-550 to factory default settings as shown in Table 4,
three confirmation tones will be heard.
After menu item 10 has been selected, cycle power to the N-550 and verify
that the factory default values have been reinstated.
.
Alarm silence restrictions
None, sound reminder, do not allow alarms off
Blip volume
0 to 10
Language
English
Serial port baud rate
2400, 9600, 19200
Serial port mode
ASCII, external equipment communications
Event in-sensor trend format
SpO
2
, SpO
2
and pulse rate
RS-232 nurse call polarity
Normally high, normally low
Table 3: Institutional Default Values (Continued)
Parameter
Range
Table 4: Factory Default Settings
Parameter
Range
Factory
Default
Setting
%SpO
2
Upper Alarm
Limit
Lower Alarm Limit plus 1 to
100%
100%
%SpO
2
Lower
Alarm Limit
20% to Upper Alarm Limit
minus 1
85%
Pulse Rate Upper
Alarm Limit
Lower Alarm Limit plus 1 to
250 bpm
170 bpm
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...