DC-3000C SERIES ULTRASONIC THICKNESS GAUGE
- 5 -
7. Preparation before measurement
【
7.1
】
Preparation of the instrument
For the newly purchased instrument, please check the instrument and its
accessory according to the “3 standard delivery”. If you find it is not the
same as the table listed, please contact the manufacture in time. If the
instrument is damaged, please do not use it and contact the manufacture
as soon as possible.
【
7.2
】
Selection of the Probe
Users can select the suitable probe according to the thickness of the
workpiece to be measured.
Type
Freq.
Meas. rang
Temp.
D 5301
5.0MHz
E-E 3.0
~
25mm
T-E 1.44
~
200mm
<60
℃
D5008
5.0MHz
0.8
~
300mm
<60
℃
D7006
7.5MHz
0.65
~
50mm
<60
℃
D7004
10.0MHz
0.65
~
20mm
<60
℃
D2012
2.0MHz
2.0
~
400mm
<60
℃
D5113
5.0MHz
2.0
~
200mm
<350
℃
Type
Application
D5301
Used in the thickness measurement through the coating
(E-E)and normal measurement (T-E)
D5008
Used common in many measurements, for example when
the measuring surface is flat or with huge curvature, or the
thickness of the workpiece is large than 50mm.
D5113
Used in the thickness measurement when the temperature
is less than 350
℃
D7006
Used in the measurement of thin wall thickness and small
curvature surface.