Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
117
11
Configuration
CAUTION
FALSE DIAGNOSIS
Wrong configuration settings can cause incorrect diagnostic calculations.
Be aware that wrong configuration settings can affect the predicted values,
the system interpretation, and the displayed result values.
Verify the configuration settings.
11.1
Configuration of
EasyOne Air
You can configure
EasyOne Air
in the tools menu.
11.1.1
Configurations in the settings menu of
EasyOne Air
>
11.1.1.1
Description of device settings of
EasyOne Air
>
>
On this screen, you can choose basic settings for
EasyOne Air
.
Setting
Purpose
Options
Default setting
Speaker volume
To set the speaker volume
0% – 100%
50%
Key click
To enable sound for audible click
feedback
Selected, deselected
Selected
Beep during
maneuver
To enable a beeping signal during
breathing maneuvers as feedback
about air flow
Selected, deselected
Selected
User Logon required
To enable password protection
Selected, deselected
Selected
System Time
To set the time
Numeric entry
(Not applicable)
System Date
To set the date
Numeric entry
(Not applicable)
Date format
To set the format of the date
YYY-MM-DD, MM/DD/
YYYY, DD.MM.YYYY
(Determined by
regional setting at
first-time setup)
Time format
To set the format of the time
24h (00:00 – 23:59), 12h
(AM/PM)
(Determined by
regional setting at
first-time setup)
Patient ID format
To set the format of the patient ID for
compliance with national standards
Numbers only,
Alphanumeric
(Determined by
regional setting at
first-time setup)
Description of device settings