Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
117
Print Lung Age
You can choose to print the lung age on reports. When the calculated lung age
is lower than the patient’s actual age, the patient’s actual age is displayed.
To display the lung age only for smokers, choose
Smoker only
from the drop-
down list. You can set a patient’s smoker status when you add patient data to
the database or edit patient data.
To always display the lung age, choose
On
.
To never display the lung age, choose
Off
.
The default configuration is
Smoker only
.
11.2.4.5
Utilities > Configuration > Test > SVC
Parameter
To select which diagnostic parameters to include on the test result screens
and on the printed reports, choose
Select FVC
and
Select FVL
.
Type
To make regular breathing mandatory before the breathing maneuver, choose
Tidal Breathing Required
.
11.2.4.6
Utilities > Configuration > Test > MVV
Parameter
To select which diagnostic parameters to include on the test result screens
and on the printed reports, choose
Select FVC
and
Select FVL
.
11.2.4.7
Utilities > Configuration > Test > CalCheck
Syringe Volume
Calibration syringes are available in various sizes. You can choose the volume
of the calibration syringe from the drop-down list.
Type Selection
One test with three trials is required for the Single Flow calibration check. For
the Multi Flow calibration check, three tests comprising three trials, each with
different flow rates, are required.
Reported Graph
You can select which type of curve you prefer for the calibration check report.