Safety information
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
16
2
Safety information
2.1
Classification
In this manual, safety information is classified as follows:
WARNING
WARNING …
… indicates a hazard. If not avoided, the hazard can result in death or serious
injury.
CAUTION
CAUTION …
… indicates a potential hazard. If not avoided, the hazard may result in minor
injury and/or product/property damage.
2.2
General safety information
Electric Shock
Patients and technicians may be exposed to dangerous voltage.
Connect only medical-grade power supply released by the manufacturer.
To ensure electric safety, connect only equipment, such as printers and net-
works, that complies with the IEC 60950-1 standard for the electrical safety of
IT equipment.
Patient health
hazard
Inadequate qualification of personnel who operate
EasyOne Air
can put the
patient’s health at risk.
The patient may suffer an asthma attack or a bronchospasm while performing
spirometry.
EasyOne Air
must only be used by qualified personnel.
Do not leave the patient unattended while performing spirometry or in be-
tween trials.
Intended audience of this Operator’s Manual, 3
Cross-
contamination and
adverse reaction
After use, the disposable
EasyOne Flow Tube
can be infectious.
Provide means for proper disposal of infectious waste, according to legal re-
quirements.
False diagnosis
Inadequate qualification of the personnel that operates
EasyOne Air
can lead
to false results, false interpretation, and false diagnosis.
EasyOne Air
must only be used by qualified personnel.
Intended audience of this Operator’s Manual, 3
Electric shock
Do not open
EasyOne Air
.