Configuration
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
117
11.2.3.4
Activating user handling and administering user accounts
1
Choose
Utilties > Configuration > General > User Handling
.
2
Enter the default user name / password and choose
Login
.
The default user name / password is:
admin
/
8005
3
To activate user handling, select the
User Handling
check box.
4
To add a user, do the following:
a)
Choose
Add
.
b)
Enter a user ID.
c)
Enter additional information as required.
d)
Choose
OK
.
5
To edit a user, do the following:
a)
Select an existing user.
b)
Edit information as required.
c)
Choose
OK
.
6
To delete a user, do the following:
a)
Select an existing user.
b)
Choose
Delete
.
c)
Confirm the following message.
7
To save the changes, choose
OK
.
8
To discard the changes, choose
Cancel
.
About user handling and the default password of EasyOne Connect, 105
11.2.4
Description of test configurations of
EasyOne Connect
11.2.4.1
Utilities > Configuration > Test > General
Best trial/best value selection
If you choose the
Best Trial
option, the complete best trial is selected for
assessing the quality of the test and for assessing the interpretation. The trial
with the largest sum of FVC and FEV
1
is selected, as suggested by ATS and
ERS.
If you choose the
Best Value
option, the best single, relevant values are
selected from different trials for assessing the quality of the test and for
assessing the interpretation. This compiled column shows, for example, the
largest FVC (or FEV
6
) and the largest FEV
1
from all acceptable tests (unless all
tests are unacceptable). Other parameters are taken from the best trial (also
defined by the largest sum of FEV
1
and FVC).