SHOULDER PACEMAKER
TM
- User Manual 04-06EN
12
Guidance and manufacturer's declaration - electromagnetic immunity
Shoulder Pacemaker
TM
is designed to operate in the electromagnetic environment specified below. The customer or user of the device should
ensure that it is used in such an environment. Portable and mobile RF communications equipment should not be used closer to any part of
the device, including cables, than the recommended separation distance calculated with the equation applicable to the transmitter
frequency.
Proof of Immunity
Test level of IEC 60601
Level of
compliance
Recommended separation
distance
d:
RF Irradiate
IEC 61000-4-3
10 V/m
da 80 MHz a 2,7 GHz
10 V/m
d= 30 cm
Proximity field
immunity from IEC
61000-4-3 RF
wireless
communication
devices
TETRA 400
380 – 390 MHz
27 V/m
27 V/m
d= 30 cm
GMRS 460
FRS 460
430 – 470 MHz
28 V/m
28 V/m
LTE Band 13, 17
704 – 787 MHz
9 V/m
9 V/m
GSM 800/900, TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
800 - 960 MHz
28 V/m
28 V/m
GSM 1800; CDMA 1900; GSM 1900; DECT;
LTE Band 5
1700 – 1990 MHz
28 V/m
28 V/m
Bluetooth, WLAN, 802.11 b/g/n, RIFD 2450, LTE
Band 70
2400 – 2570 MHz
28 V/m
28 V/m
WLAN 802.11 a/n
5100 – 5800 MHz
9 V/m
9 V/m
Table 3:
electromagnetic immunity