Chapter 16 Reference
16.5 Specifications
F-52, 53, 54, 55
245
16.5
Specifications
●
Measuring object for each model
F-52 F-53 F-54 F-55
Description
pH
JIS form
●
●
●
●
0
Measuring
principle
●
●
●
●
Glass electrode
Display range
●
●
●
●
pH -2.000 to 16.000
Measuring
range
●
●
●
●
pH 0.000 to 14.000
Resolution
●
●
●
●
0.01/0.001 pH
Repeatability
●
●
●
●
±
0.001 pH
±
1digit
mV
Measuring
range
●
●
●
●
±
1999.9 mV
Resolution
●
●
●
●
0.1 mV
Repeatability
●
●
●
●
±
0.1 mV
±
1digit
Temp. Measuring
range
●
●
●
●
0.0 to 100.0
℃
Resolution
●
●
●
●
0.1
℃
Repeatability
●
●
●
●
±
0.1
℃±
1digit
ION
Measuring
principl
●
●
Ion electrode
Measuring
range
●
●
0.00
μ
g/L to 999 g/L(mol/L)
Resolution
●
●
Valid numbers 3 digits
Repeatability
●
●
±0.5% ±1 digit of full scale
Con-
ductiv-
ity
Measuring
principle
●
●
2 AC bipolar method
Measuring
range
●
●
Cell constant
100 m
-1
:
0.000 mS/m to 19.99 S/m
Cell constant
10 m
-1
:
0.0
μ
S/m to 1.999 S/m
Cell constant
1000 m
-1
:
0.00 mS/m to 199.9 S/m
Resolution
●
●
0.05% of full scale
Repeatability
●
●
±0.5% ±1 digit of full scale
Salin-
ity
Measuring
principle
●
●
Conversion from conductivity value
Measuring
range
●
●
0.00 to 4.00%
Resolution
●
●
0.01%
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...