Chapter 15 Maintenance and troubleshooting
15.1 Maintenance
F-52, 53, 54, 55
193
15.1.2 Ion electrode maintenance
Refer to the electrode operation manual for how to take care of each kind of electrode.
●
Ion electrodes
(*)
:
Change in the electric potential of the electrode (25ºC)
when the ion concentration is changed by a factor of 10.
The above electrodes are subject to change without notice.
Maintenance of the ion electrodes listed in the above table begins, on next page.
Ion to be
measured
Ion
type
Slope
(*)
Measurement
range
Elect
-rode
mod-
el
Comp
atible
tip
model
standard
solution
Potassium
K
+
+1
+58 mV
0.04 to 39,000
mg/L
6582
-10C
7682
3.33 mol/L
NaCl
Calcium
Ca
2+
+2
+29 mV
0.4 to 40,080
mg/L
6583
-10C
7683
3.33 mol/L
KCl (#300)
Chloride
CI
-
-1
-59 mV
0.4 to 35,000
mg/L
6560
-10C
7660
1 mol/L
KNO
3
Fluoride
F
-
-1
-59 mV
0.02 to 19,000
mg/L
6561
-10C
7661
3.33 mol/L
KCl (#300)
Nitrate
NO
3
-
-1
-55 mV
0.06 to 62,000
mg/L
6581
-10C
7681
1 mol/L
KCl
Ammonia
NH
3
+1
-59 mV
0.1 to 1,000
mg/L
5002
-10C
Included
internal solu-
tion NH
4
Cl
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...