Chapter 11 METER SETTING
11.2 System setup
134
HORIBA
11.2.3 AutoHold setting
AutoHold setting is the function in which the meter judges the potential stability to assure the
measurement values. As the stability condition, five types of setting mode are available.
The AutoHold condition for calibration and inspection should become Standard in regardless
of the selected mode option.
Due to operation processing, the hold condition is not always given even when it seems to
satisfy the stabilization criteria.
Setting to “EXACT” gives the time to HOLD longer.
The AutoHold judgment for the resistivity is calculated by the conductivity values.
Mode
Measuring
target
Criteria
EXACT pH, mV, ION Potential change in 10 seconds: within 0.3 mV
Temperature change in 10 seconds: within 2.0
℃
COND
Potential change in 10 seconds: within 1digit
Temperature change in 10 seconds: within 2.0
℃
Salinity
Potential change in 10 seconds: within 0.03%
Temperature change in 10 seconds: within 2.0
℃
NOR-
MAL
pH, mV, ION Potential change in 10 seconds: within 1 mV
Temperature change in 10 seconds: within 2.0
℃
COND
Potential change in 10 seconds: within 3 digit
Temperature change in 10 seconds: within 2.0
℃
Salinity
Potential change in 10 seconds: within 0.10%
Temperature change in 10 seconds: within 2.0
℃
BRIEF
pH, mV, ION
Potential change in 10 seconds: within 3 mV
Temperature change in 10 seconds: within 2.0
℃
COND
Potential change in 10 seconds: within 5 digit
Temperature change in 10 seconds: within 2.0
℃
Salinity
Potential change in 10 seconds: within 0.3%
Temperature change in 10 seconds: within 2.0
℃
TIME
The values are held at the set time.
Setting range: 2 to 999 (sec)
OFF
The AutoHold function becomes invalid.
When set to OFF, pressing MEAS key will hold the values
at that time. (Manual HOLD)
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...