6 Maintenance and Troubleshooting
6.2 ION electrode maintenance
D-52/53/54/55
145
●
Short-term storage
Immerse electrodes in the following solutions, when they
are to be stored for up to one day and then reused.
ION electrode
Storage solution
Cl
-
ION electrode
de-ionized water
F
-
ION electrode
NO
3
-
ION electrode
1 mol/L potassium nitrate solution
(100 g/L KNO
3
)
K
+
ION electrode
0.1 mol/L potassium chloride
solution (75 g/L KCl)
Ca
2+
ION electrode
Tap water
NH
3
ammonia
electrode
0.01 mol/L ammonium chloride
solution