User Manual - Stand-Up Lift: WAYUP 5 Range

Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)

Troubleshooting Guide

Main symptom

Refined symptom

Cause

Solution

The unloaded device does not always go 

down.

The device runs empty.

Our devices need to have a weight exerted 

on them in order to be lowered.

Push down on the lifting arm while pressing 

the down button on the remote control.

The moving parts of the devices are hard, 

the device is hard to handle. 

Problem of rolling, of knee supports 

adjustment.

This is due to a lack of lubrication of the 

moving parts. 

Lubricate the moving parts.

The batteries are discharged or worn out. Charge the device or change the batteries.

The motherboard is defective.

Change the motherboard.

Battery level off while pressing the remote 

control.

Actuator cable is not connected or 

damaged.

Reconnect the actuator cable or if it is 

damaged, repair it.

Battery is charged on direct current 

(flashing battery level display) but no 

display on the control box if it is 

disconnected from the direct current.

Emergency stop button is pressed or 

defective.

Disengage the emergency stop button or 

change it.

The contact between the battery and the 

control box is not present. 

Remove and replace the battery, making 

sure that the two parts are in contact.

The battery is too worn or too discharged to 

be recharged.

Change the battery.

One of the fuses T10A (one in the control 

box and one in the battery cartridge) is cut. 

Replace the faulty fuse(s).

Battery is charged on direct current 

(blinking battery level display) and normal 

display on the control box if it is 

disconnected from the direct current.

The remote control is out of service (check 

if the device works with the buttons on the 

control box). 

Change the remote control.

In all other cases.

The motherboard is defective.

Change the motherboard.

The batteries are discharged or worn out. Charge the device or change the batteries.

The weight lifted by the device is too 

important, a weight symbol is displayed on 

the screen.

Decrease the weight lifted by the unit by 

relieving the load on the lifting arm.

The duty cycle of the actuator is reached 

(self-preservation), an actuator is displayed 

on the screen. 

Do not use the device for at least 20 

minutes.

The thermal protection of the actuator has 

been activated.

Allow the actuator to cool down.

The batteries are discharged or worn out.

Charge the device or change the battery.

The motherboard is defective.

Change the motherboard.

On direct current, the battery level flashes 

then becomes fixed in a short time (<5 

hours for a fully discharged battery).

The batteries are worn.

Change the battery.

The batteries are worn.

Change the battery.

One of the fuses T10A (one in the control 

box and one in the battery cartridge) is cut.

Replace the faulty fuse(s).

Check that the electrical outlet is functional.

Changer de prise de courant

Check if the power cord is not damaged.

Change the power cord.

One of the T1A fuses is cut

(on VEOBOX 1 and 2)

Replace the faulty fuse(s).

The internal case charger is out of service

(on VEOBOX 3).

Change the internal charger.

The motherboard is defective (on VEOBOX 

1 and 2).

Change the motherboard.

On direct cuurent, the battery level does 

not blink, it becomes fixed right away. 

The battery no longer holds the charge.

Device empty and/or being lowered 

On direct current, nothing happens.

The actuator does not work but we hear a 

"click" in the control box when pressing the 

remote control. 

The actuator does not work and there is no 

"click" sound in the control box when the 

remote control is pressed. 

Battery level remains displayed on the pack 

even while pressing on the remote. 

Battery recharged to the direct current 

(battery level display on 4 bars) but no 

display on the control box if it is 

disconnected from the direct current.

Only when the patient is hanging from the 

sling during the lifting movement. 

The actuator stops intermittently.

Summary of Contents for WAYUP 5 PLUS

Page 1: ...1 MEDICAL DEVICE CONFORM TO STANDARD NF EN ISO 10535 2007 NAUSICAA Medical is certified ISO 13485 www nausicaa medical eu User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S...

Page 2: ...l Safety Guidelines Technicals Dimensionals Characteristics WAYUP 5 WAYUP 5 avec cartement M canique des Pieds WAYUP 5 PLUS avec cartement lectrique des pieds Technicals Dimensionals Characteristics B...

Page 3: ...r a toilet or a wheelchair to be transferred while being active 1 The stand up lift has already been inspected at the factory to ensure that it is free from defects and that nothing is missing Neverth...

Page 4: ...do not overtighten Assembly is done with the brakes locked Assembling Instructions WAYUP 5 Step 1 Take the device out of the box Step 2 Retrieve the handlebar and the foot rest Step 3 Mounting the ha...

Page 5: ...s no wear or deformation on the hooks Assembling Instructions WAYUP 5 with mechanical opening legs WAYUP 5 PLUS with electrical opening legs Step 1 Take the device out of the box The device is ready f...

Page 6: ...ss the parking lever pointing upwards Castor Released When locking the castors always insure that both castors are locked If only one castor is locked on a sloping floor the unlocked wheel will roll d...

Page 7: ...m tonicity patients How to properly use a stand up lift 1 Low tonicity patients shall be transferred in a semi seated position Instructions for Use Your stand is designed to lift people so do not use...

Page 8: ...a stand up lift 2 Medium tonicity patients shall be transferred in a standing position Likewise the patient s shins must be in contact with the shin block at all times In all cases the caregiver must...

Page 9: ...status of the control box and the battery Check that the actuator is properly secured Check that the battery is charged Stand up Lifts Use 7 Control Box 2 Functions Control Box 4 Functions User Manual...

Page 10: ...concerned are WAYUP 5 WAYUP 5 with mecha nical opening legs The products concerned are WAYUP 5 WAYUP 5 with mechanical opening legs The products concerned are WAYUP 5 Plus with electrical opening legs...

Page 11: ...too much Standing up for PATIENT WITH CHEST TONICITY INSTRUCTIONS FOR USE Position the bottom of the backrest in the lumbar region in order that the ears of the attachment system are on either side of...

Page 12: ...ety An assessment of the patient capacity is required before using a stand up lift and more precisely to make sure that the patient can bear weight on at least one leg and has sufficient tonicity in t...

Page 13: ...4 66 51 50 47 Web www nausicaa medical com 12 All e du Piot ZA P le Actif 30660 Gallargues le Montueux FRANCE Tel 33 4 66 51 50 80 Fax 33 4 66 51 50 47 Web www nausicaa medical com 12 All e du Piot ZA...

Page 14: ...TENANCE A POORL Y MAINTAINED BATTERY DETERIORATES SAFETY INSTRUCTIONS BEFORE EACH USE Verify the presence and conditions of safety devices all axis and their constrained fixations pins clips and bolts...

Page 15: ...Labelling 13 For all models of the WAYUP 5 Range User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 1 5 3 4 2 7 8...

Page 16: ...ensure through adequate measures and instructions that mechanical stress of the charging cord through bending tension shear crushing is excluded during loading or cleaning of the part This also appli...

Page 17: ...height Low handles Position 1 1275 Position 2 1300 Position 3 1320 Handles high Position 1 1280 Position 2 1305 Position 3 1325 Low handles Position 1 1297 Position 2 1322 Position 3 1342 Handles high...

Page 18: ...range 320 390 rounding diameter 940 1100 total length 945 1070 total width closed feet 565 540 Maximum charge 180kg 180kg Total weight of the device 29kg 36kg Mass of the heaviest piece BASE 10kg BAS...

Page 19: ...11 cm Electronic protection when overloaded Protections types T l commande IP56 Remote control IP54 Control box IP65 Battery pack IP65 Actuator IP55 Stop limit switch Box in ABS Handcontrol with spira...

Page 20: ...d s le v rin va s arr ter et une enclume s affichera Il sera alors poursuivre le fonctionnement de l appareil ompl te e toutes les charges incompl tes de la batterie inf rieures fre correspond au nomb...

Page 21: ...steam use with a microfiber mop Do not direct steam directly on the electrical boxes MONTHLY MAINTENANCE Check that the device does not present any visible damages Check that no part is missing Check...

Page 22: ...ort check Sling fixation hooks check State of the frame and of the painting check Additional checks for models with opening base legs Foot pedal or opening legs actuator check Spreaded legs system che...

Page 23: ...sembly Pivot of wheels Sub patellar support slide ECP Pedal or Cylinder Assembly Foot Spacer System Assemblies of the right and left feet on Embase 8 Periodic lubrication Use a 3 in 1 product or simil...

Page 24: ...pport runner ECP Pedal or Cylinder Assembly Foot Spacer System Assemblies of the right and left feet on Embase 8 22 Periodic lubrication Use a 3 in 1 product or similar 4 8 5 4 8 4 3 1 2 6 6 12 12 6 4...

Page 25: ...ICKER 11 7 10 9 4 14 1 2 3 NUMBER DESIGNATION REFERENCE 0 LIFTING ARM FOR WAYUP 5 1 2 3 4 WP5 BRS PF 2 LEFT HANDLE FOR RANGE 5 LIFTING ARM WP5 BRSG PF 3 RIGHT HANDLE FOR RANGE 5 LIFTING ARM WP5 BRSD P...

Page 26: ...RIGHT HANDLE FOR RANGE 5 LIFTING ARM WP5 BRSD PF 4 LIFTING ARM GRIP HANDLES set of 2 WP PGN PF 5 MAST FOR WAYUP 5 WP5 MT PF 6 HANDLE BAR FOR WAYUP 5 WP5 GM PF 7 LEFT FOOT FOR EASYLEV AND WAYUP 5 WP5...

Page 27: ...OR RANGE 5 LIFTING ARM WP5 BRSD PF 4 LIFTING ARM GRIP HANDLES set of 2 WP PGN PF 5 HANDLE BAR FOR WAYUP 5 WP5 GM PF 6 MAST FOR WAYUP 5 WP5 MT PF 7 REMOVABLE PLATFORM FOR EASYLEV AND WAYUP 5 WP5 EP PLA...

Page 28: ...IBLE TEMPORISE 5 x 20 1A 250 v par 5 2 1 BOUTON D ARRET D URGENCE UMEC VB COMPLET 2 2 CARTER AVANT BOITIER DE CONTR LE UMEC 2 3 CARTE DE COMMANDE AVEC AFFICHEUR POUR BOITIER DE CONTR LE 2 OU 4 FONCTIO...

Page 29: ...1 8 FUSIBLE TEMPORISE 5 x 20 1A 250 v par 5 2 1 BOUTON D ARRET D URGENCE UMEC VB COMPLET 2 2 CARTER AVANT BOITIER DE CONTR LE UMEC 2 3 CARTE DE COMMANDE AVEC AFFICHEUR POUR BOITIER DE CONTR LE 2 OU 4...

Page 30: ...POUR TELECOMMANDE 2 FONCTIONS COMPATIBLE BOITIER DE CONTROLE VEOBOX III VS TLCM3 KIT AIMANT AVEC VIS ET PLAQUE DE FIXATION POUR TELECOMMANDE 2 ET 4 FONCTIONS VS AMTL FACADE POUR TELECOMMANDE 2 FONCTIO...

Page 31: ...E LEVAGE 8000N tous mod les VS VR 29 User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 4 1 6 3 2 5 NUMBER DESIGNATION REF...

Page 32: ...HAPE D EMBASE DE VERIN D ECARTEMENT DES PIEDS 3000N GRIS VS VRN ECP3 CEMB 3 CORDON SPIRALE POUR VERIN D ECARTEMENT DES PIEDS 3000N GRIS VSVRN ECP3 CRD Opening Legs Jack 3 000 Newtons NUMBER DESIGNATIO...

Page 33: ...The batteries are discharged or worn out Charge the device or change the batteries The weight lifted by the device is too important a weight symbol is displayed on the screen Decrease the weight lifte...

Page 34: ...32 User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer...

Page 35: ...does not apply if the claims are consecutive to accident misuse or neglect of the unit by the end customer shipping performed without adequate protection alteration or transformation not validated by...

Page 36: ...28 User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer...

Page 37: ...29 User Manual Stand Up Lift WAYUP 5 Range Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer...

Page 38: ...e 33 0 466 717 184 Fax 33 0 466 532 305 Email export nausicaa medical com Website www nausicaa medical eu Manufactured by NAUSICAA M dical S A S 12 All e du Piot ZA P le Actif 30660 Gallargues le Mont...

Reviews: