background image

v02.12/14

User Manual - Stand-Up Lift: MAXXI UP 300

Manufactured by NAUSICAA Médical S.A. / Approved by Ghizlane Labrosse (Biomedical Engineer)

MADE FROM ALUMINIUM

WEIGHT CAPACITY : 300 kg

CLASS 1 MEDICAL DEVICE

CONFORM TO STANDARD NF EN ISO 10535 : 2007

STAND-UP LIFT: MAXXI UP 300

www.nausicaa-medical.eu

Summary of Contents for MAXXI UP 300

Page 1: ...00 Manufactured by NAUSICAA Médical S A Approved by Ghizlane Labrosse Biomedical Engineer MADE FROM ALUMINIUM WEIGHT CAPACITY 300 kg CLASS 1 MEDICAL DEVICE CONFORM TO STANDARD NF EN ISO 10535 2007 STAND UP LIFT MAXXI UP 300 www nausicaa medical eu ...

Page 2: ...User Manual Stand Up Lift MAXXI UP 300 Manufactured by NAUSICAA Médical S A Approved by Ghizlane Labrosse Biomedical Engineer ...

Page 3: ...haracteristics Spare Parts Cleaning Maintenance Motorization Technicals Characteristics Troubleshooting Guide Warranty 2 3 4 5 6 7 8 9 10 11 12 13 User Manual Stand Up Lift MAXXI UP 300 Manufactured by NAUSICAA Médical S A Approved by Ghizlane Labrosse Biomedical Engineer Contents User Manual Stand Up Lift MAXXI UP 300 ...

Page 4: ...r or deformation on the hooks WARNING To preserve the inserts do not overtighten Assembly is done with the brakes locked Step 1 You have to find an item in the box Your stander is ready for use 1 Assembling Instructions User Manual Stand Up Lift MAXXI UP 300 Manufactured by NAUSICAA Médical S A Approved by Ghizlane Labrosse Biomedical Engineer ...

Page 5: ...3 Assembling Instructions User Manual Stand Up Lift MAXXI UP 300 Manufactured by NAUSICAA Médical S A Approved by Ghizlane Labrosse Biomedical Engineer ...

Page 6: ...lding D Handcontrol status Preventive Maintenance Security Controls Stand Up Lift MAXXI UP 300 Serial Number Control of security points Frequency depending on use recommended every three months This is a sight control any wear trace must lead to a replacement GOOD WRONG DATE OF CHANGE Control of electrical functions Frequency depending on use at least once a year this control must be carried out w...

Page 7: ... mast handle Periodical Greasing WARNING DO NOT USE THE DEVICE WITHOUT THE SHIN BLOCK THESE CONTROLS CAN BE CARRIED OUT WITHOUT LOAD 9 OVERALL STATE OF THE DEVICE WELDS JOINTS OXIDATION Preventive Maintenance Security Controls Control of security points Frequency depending on use recommended every three months This is a sight control any wear trace must lead to a replacement User Manual Stand Up L...

Page 8: ...r to ensure through adequate measures and instructions that mechanical stress of the charging cord through bending tension shear crushing is excluded during loading or cleaning of the part This also applies to the electrical cables of other devices used with the Stand Up Lift 11 Respect the activation duration and the maximum weight capacity These values must not be exceeded otherwise the Stand Up...

Page 9: ...al Height 470 l Minimal Distance Between the Wall and the SCP at its Maximal Reach 210 m Minimal Distance Between the Wall and the SCP at its Maximal Height 190 n Maximal Diameter 1300 i j m l k c d e 2 3 1 a g f n r b SCP Suspension central point Technicals Dimensionals Characteristics Frame made in aluminium Laser cutting and bending sheet metal tubes and profiles Epoxy baked painting Ball beari...

Page 10: ...E UMEC VB VS BCAM FAE FACADE ADHESIVE ECRAN CARTOUCHE BATTERIE VS BCAM VS BCAM CR CROCHET D ATTACHE CARTOUCHE BATTERIE VSBCC ECP BOITIER DE CONTRÔLE UMEC VB 4 FONCTIONS sans cartouche batterie VS BCC CCC CARTE DE COMMANDE AVEC CHARGEUR POUR BOITIER DE CONTRÔLE UMEC VB 2 ET 4 FONCTIONS VS BCC BA ECP CARTE DE COMMANDE AVEC AFFICHEUR POUR BOITIER DE CONTRÔLE UMEC VB 4 FONCTIONS VS KCNF KIT CONNEXION ...

Page 11: ...infectant solution surface Maintenance by specific service providers after removal of the stand up lift from the facility Operation biological cleaning Steam cleaning of various flat surfaces Change surfaces regularly when washing to avoid water load Steam clea ning of inaccessible areas For tubes steam use with a microfiber mop Do not direct steam directly on the electrical boxes CAUTION Disinfec...

Page 12: ...N VDE 0700t238 1983 DC motor low voltage 24V 24V 120VA power Maximum strength 10 000 N Lifting stroke 29 cm Electronic protection when overloaded Protection class II Protection type IPX 4 Stop limit switch Box in ABS Handcontrol with spiral cable low voltage 24V Protection against overloading through thermo switch Loudness less than 55dB at a distance of 1 meter Electromagnetic compatibility confo...

Page 13: ... need to be checked by the manufacturer After 10 000 complete cycles a key appears on the screen and the actuator must be checked Motorization Technicals Characteristics Overload protection system If you attempt to lift a person weighing beyond the maximum limit of the device an anvil with kg written will appear asking you to immediately relieve the unit because it is a non conform use Preservatio...

Page 14: ...oes not hear a click in the control box when the remote control button is pushed 1 Emergency button is activated 2 There is no contact between the battery and the control box 3 Batteries are unloaded 4 The remote control jack is not plugged 5 The remote control jack is deteriorated 6 There are damages on the remote control or on the control box 1 T urn the button to unlock the device 2 Take off th...

Page 15: ...s not apply if the claims are consecutive to accident misuse or neglect of the unit by the end customer shipping performed without adequate protection alteration or transformation not validated by NAUSICAA Médical S A the impact of external elements natural disaster fire shock humidity flood lightning installation and or use in a non compliant way with technical standards and safety in case the un...

Page 16: ...one 33 0 466 717 184 Fax 33 0 466 532 305 Email export nausicaa medical com Website www nausicaa medical eu Manufactured by NAUSICAA Médical S A 12 allée du Piot ZA Pôle Actif 30660 GALLARGUES LE MONTUEUX FRANCE ...

Reviews: