Xltek® Trex HD Amplifier
User & Service Manual
58
15. Maintenance, Cleaning, & Disposal
To keep the Trex HD amplifier in good working condition, follow a regular schedule of maintenance.
Regular maintenance performed by the user does not involve access to the interior of the Trex HD
amplifier and components. For servicing problems that require corrective maintenance and/or internal
component service, call Technical Support at
1-800-303-0306
, or contact your local
Natus
representative.
Periodically check cable connections and electrodes for damage and wear. Inspect cables for bent pins.
Replace frayed or worn cables. Also, regularly inspect and clean all system components, including:
Connectors and jack ports
Amplifier and USB 2.0 cable
Electrodes and accessories
The Trex HD amplifier and its components should not be immersed in water or any other fluid. To clean,
use a damp cloth to wipe all surfaces.
The Trex HD amplifier is designed to be extremely portable. As such, the Trex HD is subject to increased
daily wear and tear. Taking basic care of the system and avoiding extreme physical abuse helps prolong
the lifespan of the Trex HD.
15.1. Recommendations
Disconnect all cables from the Trex HD before wiping. Wipe using a soft, damp cloth using
non-conductive distilled water, electrically non-conductive inert surfactants or an
Natus
-
approved cold sterilizing agent. Do not use any abrasive cleaner on the system.
It is important to dry off the unit quickly. Be careful not to allow any fluid to seep into the
internal electronic components of the system.
Do NOT leave the amplifier attached to the computer when transporting the unit.
Do NOT autoclave, pressure sterilize, or gas sterilize this amplifier.
Do NOT soak or immerse the amplifier in any liquid.
A cleaning solution of 70% isopropyl alcohol is recommended.
Use cleaning solution sparingly. Excessive solution can flow into the amplifier and cause
damage to internal components.