Page 1: ...Natus Photic Stimulator User Service Manual...

Page 2: ......

Page 3: ...use extend to all media in which this information is preserved This copy of the User Manual shall be used only in accordance with the conditions of sale of Natus Neurology Incorporated or its distribu...

Page 4: ...ENDED USER PERFORMED MAINTENANCE 10 Cleaning 10 Inspection 10 Maintenance 10 ENVIRONMENTAL SPECIFICATIONS 11 Temperature 11 Humidity 11 Condensation 11 Flammability 11 EMC Specifications 11 Safety Sta...

Page 5: ...by the Natus NeuroWorks TM software from laptop or desktop computers it supplies intense flashes of light Its components consist of an arm mounted light assembly a light source an intensity control a...

Page 6: ...presence of flammable anesthetics Reliable grounding requires hospital grade receptacles and power cord Always perform a leakage current test and compare to allowable standards BEFORE connecting the p...

Page 7: ...test In addition the accompanying EEG Electroencephalograph amplifier s signals will also be contaminated past the point where any clinical signal interpretation is possible Trained electroencephalog...

Page 8: ...on of warnings and markings in a restricted space ATTENTION Consult Accompanying Documents Consult Accompanying Documents Protective Earth Ground Type B equipment Type BF Equipment Dangerous Voltage A...

Page 9: ...ort arm 52 in 1320 8 mm total Frequency of flash Maximum 60 Hz Light intensity Adjustable via Intensity control 12 position control 11 settings 1 Off Input requirements TTL Positive Pulse 100 s 1 mA M...

Page 10: ...Stimulator Netlink Cable 580 PHTCB3 Photic Stimulator XL Cable Installation and Operation WARNING Never place the photic device on the floor 1 Place the photic device on a platform table cart or other...

Page 11: ...onnector 8 On the Gamma headbox connect the phone plug of trigger cable BLSC part number 580 PHTCB2 labeled Strobe Out to the phone jack on back panel of computer labeled Strobe 9 Use the photic devic...

Page 12: ...pecial cleaning is required for this device Ordinary house keeping type cleaning to prevent the accumulation of dust or debris is adequate Inspection No special inspection procedures are required for...

Page 13: ...ability UL94 HB UL94V O EMC Specifications IEC 60601 1 2 2007 Electromagnetic Compatibility EMC Testing CISPR 11 B Conducted and CISPR 11A Radiated Emissions EN 61000 4 2 Electrostatic Discharge Test...

Page 14: ...anual Room Topography Place all equipment on an even level surface Avoid the potential for mechanical shock or possible vibrations during setup system operation or when relocating the equipment Verify...

Page 15: ...ients are defined as short bursts of energy that are propagated through the power cord The EFT source is usually located in the nearby equipment or machinery such as motors switching equipment light s...

Page 16: ...rence in nearby electronic equipment RF emissions CISPR 11 Class A The Natus Photic Stimulator is suitable for use in all establishments other than domestic and may be used in domestic establishments...

Page 17: ...ment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltag...

Page 18: ...electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the fol...

Page 19: ...mulator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Natus Photic Stimulator as recommen...

Page 20: ...Natus Photic Stimulator User Service Manual 18 Intentionally Left Blank...

Page 21: ...User Service Manual Natus Photic Stimulator 19 Intentionally Left Blank...

Page 22: ...Natus Photic Stimulator User Service Manual 20 Intentionally Left Blank...

Page 23: ...Intentionally Left Blank...

Page 24: ...t is Natus Medical Incorporated an internationally respected innovator of medical products and services Our Neurology systems are backed up by an in house support team staffed with technical and clini...