OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
6-9
Repair Procedures
Figure 6.3
Disconnecting the heater control relay
5
Remove and save the two screws that secure the relay to the heatsink, then
remove the old relay.
To install the new heater relay:
1
Coat the contact surface of the heatsink with a thin, uniform layer of heatsink
compound.
2
Place the new relay with terminals 2 and 4 toward the front of the device (for
Model 43, see wiring diagram 700199; for Model 44, see wiring diagram
700203).
3
Attach the wires to the heater control relay (see page A-1).
4
Insert the front cover, then tighten the screws just inside the Drier door.
5
Plug in the power cord, then press the power switch to the
ON
(|) position.
6
Test the device for proper operation (see page 2-4).
Heatsink
Connector
bracket
Screw
Star
washers
Wire #28 at
terminal #4
Wire #25 at
terminal #3
Screws (x2)
secure relay to
heatsink
Wire #33
(Model 44 only)
Wire #9 at
terminal #1
Wire #11 at
terminal #2
Double wire
at terminal
#2 (Model
44 only)
NOTE
#
Ensure that the wires are placed on the relay terminals per the wiring diagrams and that
washers and screws are arranged (see Figure 6.3).