Nicolet Cortical Stimulator
3-24
Revised 11/21/18
Recommended workflow
In cases where it is desired to
Start
and
Stop
the stimulation directly from the
Cortical Stimulator device, perform the following steps:
1. Select
Semi-automatic
mode from the Relays control mode dialog, and
click
OK
.
2. Select
electrode pairs
for stimulation.
3. Set the
stimulation parameters
as desired.
4. Click the
Close Relays
button.
5. Use
Start
and, as needed,
Stop
on the stimulator device.
In this mode, stimulation parameters can be adjusted on the cortical
stimulator itself, parameters will be sent to the study when used.
6. Select the radio button next to the desired mode:
Automatic
,
Semi-
automatic
,
Semi-automatic plus
, or
Manual.
7. Click
OK
to close the dialog.
Mode
Use Case
Operating Mode
Automatic
mode
Used in case of full
Cortical stimulation
procedure from
NeuroWorks software,
not from Nicolet Cortical
stimulator device
DSM relays are closed
automatically prior to starting
stimulation and opened after
stimulation end.
In this mode, buttons on
stimulator are deactivated to
prevent starting/stopping
stimulation from the device.
This is because when pressing
the
Start
button on the
stimulator, the relays in
Quantum would close slightly
too late with potential missed
stimulation pulses.
Semi-
Automatic
mode
Used if closing DSM
relays is done from
NeuroWorks software but
stimulation is started from
the Nicolet Cortical
stimulator device
DSM relays are opened
automatically from
NeuroWorks software after
stimulation end.
Summary of Contents for Nicolet
Page 2: ......
Page 6: ...Nicolet Cortical Stimulator 4 Revised 11 21 18 Blank page ...
Page 11: ...Revised 11 21 18 1 1 1 General Information ...
Page 12: ...Nicolet Cortical Stimulator 1 2 Revised 11 21 18 Blank Page ...
Page 23: ...Revised 11 21 18 2 1 2 Nicolet Cortical Stimulator Overview ...
Page 24: ...Nicolet Cortical Stimulator 2 2 Revised 11 21 18 Blank Page ...
Page 41: ...Revised 11 21 18 2 19 Overview Blank page ...
Page 42: ...Nicolet Cortical Stimulator 2 20 Revised 11 21 18 ...
Page 43: ...Revised 11 21 18 3 1 3 Performing a Cortical Stimulation Study ...
Page 44: ...Nicolet Cortical Stimulator 3 2 Revised 11 21 18 Blank Page ...
Page 69: ...Revised 11 21 18 4 1 4 ElectromagneticCompatibility EMC ...
Page 70: ...Nicolet Cortical Stimulator 4 2 Revised 11 21 18 Blank Page ...
Page 77: ...Revised 11 21 18 5 1 5 Error Codes ...
Page 78: ...Nicolet Cortical Stimulator 5 2 Revised 11 21 18 Blank Page ...