Nicolet Cortical Stimulator
2-6
Revised 8/27/13
3
Bottom View
LEGEND
LEGEND
A
Status
Shows current status of the stimulator:
Set-up
,
Ready
,
Stim On
.
B
Setup
1, 2
Changes the Stimulus Control Unit (SCU) parameters.
Change Mode
Probe Biphasic
Electrode Biphasic
Select Label Set
Numeric
Montage
Delete Label Set
Select Language
A list of available languages appears
Adjust Contrast
Rotate the Selector knob for the best display contrast
C
Pulse Frequency
1, 2
The rate at which pulses are delivered (50 Hz typical). You can also select
Trigger
for external
control.
D
Pulse Duration
1, 2
The length of time for each pulse. The actual pulse length is twice the pulse duration:
100 -
1000 uSec
.
E
Train Duration
1, 2
Press to view the present maximum stimulus duration (5 seconds typical). You can also select
a
Single
train duration, or
Trigger
for external trigger control.
F
G
Ele
Electrode -
1, 2
1, 2
Selects which electrode (channel) will receive the stimulus. Rotate the
Selector Knob
to step
through the channel numbers.
H
Set Stim
1, 2
Sets the
current level
applied to the patient, measured from baseline to peak (8 mAmps or
less typical). Enter the most significant value using the
Selector Knob
. Press the
Set Stim
button
to select the least significant value and then enter by using the
Selector
knob. Repeat as
necessary.
I
Delivered Stim
Displays the
stimulation level
being delivered. The LED indicates when stimulation is being
delivered.
J
Check Stim
1
Applies the selected stimulus to an internal load to verify correct operation. The LED lights
when this feature is enabled. The actual current that is delivered is displayed in the
Delivered
Stim
display field.
K
Mark Channel
Press the
Mark
Channel
button to mark the channels selected.
L
Start
1
Delivers the stimulation pulse (
Train
,
Continuous
, or
Single
) and the external
Trigger In
signal is not being used. Must be in the
Ready
state to function.
M
Ictal Disrupt
1
Repeats the first pulse in the train.
N
Selector Knob
Turn to scroll through options, press to select.
O
Stop
Interrupts stimulation.
P
Trigger Out
Future use.
Q
Trigger In
Future use.
R
Trigger Synchronization
Future use.
S
Serial Interface
For Nicolet system interface.
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...