Natus® Brain Monitor &
Embla Dx Series™
User and Service Manual
16
The Brain Monitor & Embla Dx Series Amplifiers need special precautions regarding
Electromagnetic Compatibility (EMC) and must be installed and operated according to EMC
guidelines. Refer to the
EMC Standard of Compliance and Normative References
Using the Brain Monitor & Embla Dx Series Amplifiers with cables and accessories not approved
by Natus may negatively affect EMC performance, including electromagnetic immunity. Refer to
the
External equipment may interfere with the performance of the Brain Monitor & Embla Dx Series
Amplifiers, even if other equipment complies with CISPR Emission requirements. Refer to the
EMC Standard of Compliance and Normative References
The Brain Monitor & Embla Dx Series Amplifiers should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the amplifier should be observed to
verify normal operation in the configuration in which it will be used. Refer to the
recommended separation distances.
Do not connect items which are not specified as part of the Brain Monitor & Embla Dx Series
Amplifier system to the system.
Electrical Warnings and Cautions
Natus
systems are intended for connection to a properly grounded electrical outlet only.
Conductive parts of electrodes and their connectors are not to contact other conductive parts
and earth.
Do not place MULTIPLE PORTABLE SOCKET-OUTLETS (MPSOs) on the floor.
Do not connect additional MPSOs or extension cords to the Brain Monitor & Embla Dx Series
Amplifier systems.
ELECTRICAL SHOCK HAZARD:
Do
NOT
turn on the system power until all cables have been
connected, verified and visually inspected for any damage. Failure to inspect the cables may
result in electrocution. Verification of electrical safety should be performed routinely.
ELECTRICAL SHOCK HAZARD:
Do
NOT
service the system. Refer servicing to qualified
personnel only. Do
NOT
use repaired components without proper testing.
Do not use the MPSO with the Brain Monitor & Embla Dx Series Amplifiers system for supplying
power to any equipment that is not part of the system.
To avoid the possible hazards caused by the summation of leakage currents when all the parts
of the system are interconnected, no equipment other than devices connected to the Brain
Monitor & Embla Dx Series Amplifier systems may be powered by the isolation transformer.