Natus® Brain Monitor &
Embla Dx Series™
User and Service Manual
46
Maintenance, Cleaning, & Disposal
To keep the Brain Monitor & Embla Dx Series Amplifiers system in good working condition, follow a
regular schedule of user performed maintenance. Regular maintenance performed by the user does not
involve access to the interior of the amplifier and components. For service problems that require
corrective maintenance and/or internal component service, call Natus Technical Support at
1-800-303-
0306
, or contact your local Natus representative.
Periodically check cable connections and electrodes for damage and wear. Inspect cables for bent pins.
Replace frayed or worn cables. Also, regularly inspect and clean all system components, including:
Connectors and jack ports
Base Unit, Breakout box, and Ethernet cable
Electrodes and accessories
The Brain Monitor & Embla Dx Series Amplifiers and its components should not be immersed in water or
any other fluid. To clean, use a damp cloth or cleaning/disinfecting products such as isopropyl alcohol,
PDI “Sani Wipes AF3”, PDI “Sani-Cloth Plus”, Metrex CaviWipes, or AIC Wedge Wipes to wipe all
surfaces, cables and accessories.
Brain Monitor & Embla Dx Series Amplifiers Pouch
The pouch can be cleaned as follows:
1.
Hand wash with warm water or machine
wash on “Gentle” cycle with mild detergent.
2.
Do not bleach.
3.
Tumble dry on low or air dry.
Recommendations
Disconnect all cables from the amplifier before cleaning. Use a lint-free cloth. Do not use
cleaners on any system component.
Take care not to allow any fluid to seep into the internal electronic components of the system.
Do NOT autoclave, pressure sterilize, or gas sterilize this amplifier.
Do NOT soak or immerse the amplifier in any liquid.
A cleaning solution of 70% isopropyl alcohol is recommended.
Use cleaning solution sparingly. Excessive solution can flow into the amplifier and cause
damage to internal components.
Do NOT use petroleum-based or acetone solutions, or other harsh solvents, to clean the
amplifier.