This section provides important information about using radiometers to measure the light 
intensity output of the neoBLUE®mini LED Phototherapy device.   

 
 

Natus recommends that the neoBLUE mini light intensity output be checked with a radiometer per 
hospital protocol or at least every six months.  In addition, your radiometer should be calibrated according 
to the manufacturer’s specifications (typically on an annual basis).  Checking intensity before each use 
will confirm the “dosage” of light being delivered.  In addition, by checking the intensity at the center of the 
infant and at the infant’s head and feet, you can confirm proper placement of the light. 

 
 
The intensity of the light is inversely related to the distance from the source.   The neoBLUE mini LED 
Phototherapy system can be placed closer to the infant, yielding a higher intensity and thus delivering a 
higher dosage treatment.  Please refer to Section 2.3.2, Adjusting Light Intensity, to adjust the neoBLUE 
mini light to the desired intensity when changing the distance.  Natus recommends calibrating to between 
30 – 35 

μ

W/cm

2

/nm. 

 
 
The light intensity output of the neoBLUE mini device is measured with the neoBLUE® Radiometer by the 
manufacturer prior to shipment.  Because your facility may use a different radiometer to measure light 
intensity output, it is necessary to understand how your reading correlates to the Natus reading.  The 
following table provides a light intensity output conversion chart, when measured at 12 inches (30.5 cm) 
from the center of the neoBLUE mini light enclosure.  

 
NEOBLUE LIGHT INTENSITY OUTPUT CONVERSTION CHART 

Radiometer: 

Natus® 
neoBLUE

® 

Radiometer * 

Ohmeda Bili- 
blanket® Meter 
II 

Joey® 
Dosimeter 
Model JD100 

Olympic® Bili- 
meter® 

35 35 91 23 

Intensity: 
(µW/cm

2

/nm) 

30 30 78 20 

 

* This Radiometer (P/N 53870) is used for measuring the neoBLUE light intensity output prior to shipment. 

 

If you have any questions, please contact Natus Medical Incorporated Technical Service at 1-800-
303-0306 or your authorized service representative. 

 

Summary of Contents for neoBLUE mini

Page 1: ...SERVICE MANUAL neoBLUE mini System Service Manual P N 051466F...

Page 2: ...printed or copied in whole or in part without written consent from Natus Medical Incorporated The content of this manual may change without notice Natus Medical Incorporated 5900 First Avenue South Se...

Page 3: ...3 2 Adjusting Light Intensity 3 Troubleshooting the neoBLUE mini light 4 Repair Procedures 4 1 Removing the Light Lens 4 2 Removing the LED PCB 4 3 Removing the Current PCB 4 4 Removing the ON Standby...

Page 4: ...sinets or radiant warmers The arm is designed to attach to poles with 0 75 to 1 5 inch 1 91 to 3 81 cm diameters The light output is optimized to provide a peak intensity of 30 35 W cm2 nm Blue LEDs e...

Page 5: ...power on Off stand by Attention consult product documentation Always protect the baby s eyes with eye patches or equivalent Authorized European Representative XXXX Manufacture Date Natus Manufacturer...

Page 6: ...n 2 1 1 Cleaning the Interior of the Light Enclosure Section 2 1 2 Warning Disconnect light from AC power before cleaning 2 1 1 Cleaning the Light Enclosure and Adjustable Arm Remove dust from the ext...

Page 7: ...tution and country Test equipment and its application must comply with the applicable standard TABLE 2 1 Power Cord Ground Integrity TABLE 2 2 Earth Leakage Current Maximum Current Limits are listed b...

Page 8: ...locked 4 Connect the AC power cord to the UUT and to an AC receptacle Warning Operator Safety Sensitive individuals may experience headache nausea or mild vertigo if he she stays too long in the illum...

Page 9: ...not within range note the value for use later in this procedure 13 If all previous steps generate satisfactory results the UUT performance has been verified and the UUT can be returned to the user Swi...

Page 10: ...ng s goes out Multiple LED strings go out cannot attain appropriate intensity Light does not hold position Light intensity will not adjust When opening the device be sure to follow the procedures desc...

Page 11: ...o the device socket If the device will still not turn on try another power cord If none of the above restores the light open the main enclosure and check for the following Damaged power cable Check th...

Page 12: ...e infant LED STRING S GOES OUT LED short or open Increase the intensity if necessary If the device still produces enough light there is no need to replace the panel If the device does not produce enou...

Page 13: ...ny repair is completed and the UUR is reassembled The UUR can be disassembled down to all major component parts including 4 1 Removing the Light Lens 4 2 Removing the LED PCB 4 3 Removing the Current...

Page 14: ...e replacing the lens assembly obtain the replacement and reverse the previous steps 6 If no further repair is required return to chapter 2 and perform Safety and Performance Verification Tests If furt...

Page 15: ...rom the Current PCB Note The Current PCB is held in the light enclosure by positioning guides which may also come out when the LED PCB is removed 2 If you are replacing the LED PCB obtain the replacem...

Page 16: ...osest to the LED PCB 4 Remove the ON Standby switch wires by pulling them off the switch Note Polarity is not important for the switch wires However position is important Mark the two appropriate swit...

Page 17: ...previous steps in chapter 4 return to chapter 2 and perform Safety and Performance Verification Tests If further disassembly is required store the switch in a safe anti static environment and proceed...

Page 18: ...arm prior to removal Use the large Phillips screwdriver to remove the two screws that attach to the clamp Depending on the configuration of your neoBLUE mini arm the screws may be located at the front...

Page 19: ...Supply 1 2 040821 Enclosure Polycarbonate 1 3 040783 Current PCB 1 4 001839 LED PCB 1 5 800240 Nylon Washer 2 6 600122 On Standby Switch 1 7 030745 Plate w Insert 1 8 030748 Tab Washer 1 9 800244 Lock...

Page 20: ...and thus delivering a higher dosage treatment Please refer to Section 2 3 2 Adjusting Light Intensity to adjust the neoBLUE mini light to the desired intensity when changing the distance Natus recomme...

Reviews: