neoBLUE LED PHOTOTHERAPY RADIOMETER Instruction Manual
3
This instrument is calibrated at 72° F (22° C). Readings may change with
temperature and increase at the rate of approximately 0.15% per °F (0.27% per °C).
Readings are not affected by momentary exposure (3 minutes or less) to
temperature changes of less than 20° C, such as readings taken in an incubator.
Operation
Connecting the Sensor
The sensor connects to the socket on the readout. To connect, turn the connector
until the keyed slots are aligned, and then carefully insert the connector into the
socket.
Figure 3
Connecting the sensor to the readout
WARNING
Explosion hazard. Do not use this device in the presence of flammables (for example, oxygen,
nitrous oxide, and anesthetics).
CAUTION
Read and be familiar with this instruction manual before using this device.
This product should only be used by technically qualified hospital personnel.
Inspect this device before each use to ensure proper functioning.
If meter or sensor is dropped a distance of 1 foot or more, verify proper operation and compare
reading to that of a known good meter. Have unit recalibrated as necessary.
NOTE
The hospital or facility is responsible for ensuring that all personnel who operate or maintain
this device are trained in its operation and safe use, and for maintaining training records of
attendance and evidence of understanding.
Readout
Socket
Connector