Wall-mount installation
It is recommended that you connect the external power sup-
ply and the accessories before you mount Madsen A450 on
the wall.
1. Select two suitably sized screws that will pass through the
wall-mount holes on the back of the device:
Max. screw diameter 4.3 mm (0.15 in).
Max. screw head diameter: 9 mm (0.35 in).
2. The distance between the two wall-mount holes on the
back of the device is 24 cm (9.4 inches), measured from
the center of each hole.
3. Mark up the two holes on the wall and make sure that the
device will be placed horizontally.
4. Fix the two screws in the wall.
5. Hang Madsen A450 on the screws.
Madsen A450 cabling
A.
External power supply cable
B.
USB cable between Madsen A450 and the PC
5
Connecting accessories to Madsen A450
The installation must be carried out in accordance with IEC 60601-1 (Ed. 3.1), UL60601-1, and CAN/CSA C22.2
No. 60601-1-14:2014.
It is a general rule for all electrical equipment used in the proximity of the client that:
•
The connected equipment must comply with IEC 60601-1 (Ed. 3.1) except for the PC.
See also
For a detailed description of the connection panel, see the Madsen A450 Reference Manual.
Connection panel - Madsen A450
Connect the plugs to the sockets in the connection panel.
Madsen A450
7
5 Connecting accessories to Madsen A450