12
ICS Aircal
Getting started
Turn off the device and unplug from the electrical wall outlet.
Remove the internal cover.
Check if the water reservoir is full. If not, place the funnel in the refill hole of the reservoir and add additional water.
Remove the funnel and replace the internal cover. Tighten the screws to secure.
Replace the external cover and tighten screws to secure.
Plug the power cable of the device into the electrical wall outlet and turn it on.
Priming with Bent Tube
Take the syringe with the bent tube and remove enough water from the reservoir to fill the syringe.
Put the end of the bent tube into the pump inlet hole. The tube will be facing toward the front of the irrigator.
While pressing the priming switch , inject the water into the inlet.
Release the priming switch.
Press the priming switch to turn on and hold 5 seconds.
Release the priming switch to turn off and wait 5 seconds.
Repeat steps
for a total of 10 times on and 10 times off.
Press the cool button
. Decrease the temperature to 12°C.
If the temperature reaches the set temperature, then the irrigator is primed. Continue to step
Caution
• If the temperature does not reach 12°C in approximately 3 minutes or if the over temperature
alarm sounds, skip to the procedure
Note
• The default airflow rate for caloric irrigation is 8 liters per minute. If the device is transported, the airflow
knob setting might be changed. You may have to reset the airflow.