Reference Electrode Indicator
-
Red
The Reference electrode corresponds to the red input
connector.
The green light (LED) indicates the results of the impedance
test. –See the LED Status and Results description under the
Active Electrode Indicator above.
Amplifier Input Connectors (1-3) or (1-4)
The amplifier input connectors 1-3 feature both a DIN-type
socket and a pair of 1.5 mm touch-proof connectors.
Amplifier Input Connectors (4-8)
The amplifier input connectors 4-8 feature a pair of 1.5 mm
touch-proof connectors. These amplifier input connectors
can share the same reference pin.
Interconnected Reference Line / Indicator
The green light (LED) indicates that the reference input
connectors above the reference line are interconnected. This
feature is active when selected in the software application
setup.
Patient Ground Connector / Indicator
Connect the patient's ground electrode to the green
connector.
The green light (LED) indicates the results of the impedance
test. –See the LED Status and Results description under the
Active Electrode Indicator above.
WARNING
Do not connect the "patient ground" to the
protective earth connection on the Isolating Transformer, or
to any other "ground" connections, as the electrode inputs
are galvanically isolated.
Patient Ground Connector / Indicator
Connect the patient's ground electrode to the green
connector.
The green light (LED) indicates the results of the impedance
test. –See the LED Status and Results description under the
Active Electrode Indicator above.