79
List of Priorities for Each Alarm
Notice
If any of the software, firmware, or electrical components inside the monitor suffers a serious malfunction during a physiological
alarm, the Camino ICP Monitor will activate an irreversible technical alarm that will override the physiological alarm and indicate a
system failure. If this irreversible technical alarm is activated, the monitor will no longer report patient values and the user will need to
contact Natus Neuro for service. For more information on this error condition, see Chapter 5.
The following table lists the priorities of each alarm condition:
Alarm
Priority
Type of Error Condition
Type of Clinical Priority
1
Irreversible System Failures)
Medium
2
High ICP alarm
Medium
3
ICP catheter failure
Medium
4
Temperature sensor failure
Medium
5
Low battery level (5 minutes or less)
Medium
6
Monitor overheating
Medium
7
Cooling fan failure
Low
8
Mean ICP is out of accuracy range
Low
9
Temperature range is out of accuracy
Low
10
Battery failure
Low
11
Low Battery level (15 minutes or less)
N/A
12
Multiple catheters are connected
N/A
13
Catheter initializing
N/A
14
No catheter is connected
N/A
15
Catheter initialization failure
N/A
16
Cannot write to log file
N/A
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...