natus ALGO 3i Quick Reference Manual Download Page 11

Procedures, section “Printing (or Exporting) a Record Immediately after 
Screening” in the user manual.

 

Step 11. 

Power off:

 Turn off the device by sliding and holding the 

ON/OFF

 switch for 3 

seconds. The ALGO 3i screener will count down from 5 to 1 before powering off. 

Step 12. 

Clean Preamp/PCA and ATA cables before and/or after use: 

Clean patient 

cables after use. Residual sensor adhesive may corrode Preamplifier/PCA cable clips. 

SpeedScreen mode 

(alternative to

 Steps 3 to 6 

of ScreenBaby mode) 

When enabled (see Biomed preferences), SpeedScreen mode allows advanced users to 
immediately begin screening according to Biomed-selected default parameters. Infants 
should be prepped, and screening supplies/cable attachments should be in place prior to 
clicking on SpeedScreen. You may only enter infant data by clicking the 

OK

 key (to 

access Infant Data window) prior to completion of the screen. SpeedScreen may decrease 
total screening time by allowing users to enter infant data while the screening is in 
progress. 

a.  Prepare infant for screening.  

Refer to Chapter VI: Baby Preparation.

 

b.  Turn on the ALGO 3i screener, wait for the MainMenu to appear, and click on 

SpeedScreen. (If required fields are set, SpeedScreen will not be available.) Click the 

OK

 key. 

c.  If impedances are below 8 kOhms, screening will immediately begin, and ALGO 3i 

screener will advance to the “Screening in Progress” window. 
i.  If impedances are between 8 kOhms and 11 kOhms, the “Impedances Marginal” 

message will appear, showing actual impedances, and asking you to confirm your 
intention to begin screening. Natus recommends skin prepping to reduce 
impedances to below 8 kOhms. 

ii.  If impedances are 12 kOhms or greater, the “Impedances Too High” message 

will appear and screening will pause. Skin prepping is necessary. 

iii.  Once impedances are satisfactory, click the 

OK

 key to begin or continue 

screening. 

iv.  The ALGO 3i screener will advance to the Screening in Progress window. 

d.  If you wish to enter infant data for the screening record, click 

OK 

from the Screening 

in Progress window to access the Infant Data window.  
i.  You must access the Infant Data window prior to completion of the screen 

(minimum time is about 36 seconds), or else the screening record will be saved 
without any infant data. 

ii.  The device will display the Infant Data window. Enter the desired information, 

and click 

OK 

to return to the Screening in Progress window. As in ScreenBaby 

mode, the ALGO 3i screener will display the information you have entered and 
will ask you to confirm that all data is correct. 

Note:

 There is no time constraint for entering infant data once you are in the 

Infant Data window. The screen may be completed while you are entering infant 
information but the ALGO 3i device will wait until you click 

OK

 before matching 

this infant information to the screening results and save it as a screening record. 

Natus Medical • www.natus.com 

Page 11 of 12 

P/N 051179E 

Summary of Contents for ALGO 3i

Page 1: ...h filter settings Note Ensure that the notch filter setting of 50Hz or 60Hz is correctly adjusted for your region An incorrect setting may result in high myogenic interference while screening 6 Set th...

Page 2: ...OK Key Up Down Arrow Keys The display will always indicate to the user which keys are used to select scroll advance to the next window or go back to the previous window Examples Go Back to MainMenu Ad...

Page 3: ...e at a time Multi choice options are indicated by a toggle key symbol See illustration below In this case the black select bar indicates use of the toggle keys to select or deselect the desired option...

Page 4: ...e black select bar will indicate which data field is active and a cursor symbol will indicate the current position within the data field To begin scrolling through the keypad click on the up or down t...

Page 5: ...up Warnings The ALGO 3i screener will briefly display a Start Up window while it performs a self check Warnings may appear which may require your acknowledgement The MainMenu will appear next Step 2...

Page 6: ...There are three data fields in the first Infant Data Window Infant Identifier Last Name and Frist Name There are another three data fields in the second window Date of Birth User Identifier and Comme...

Page 7: ...back Common site front or back of either shoulder not on cheek Note Ensure the tab portion of all sensors are pointed towards the same side of baby Note Hold sensors in place for a few seconds to war...

Page 8: ...proceed 35 dBnHL screening only default setting 40 dBnHL screening option enabled Step 6 Start Screening Check Impedances Test immediately begins after screening parameters are selected Monitor the i...

Page 9: ...reening in progress Monitor impedances and the Message window during the screening The ALGO 3i screener will alert you to any conditions which require your attention during the screening process The S...

Page 10: ...nnected Battery charger connected Screening time limit exceeded Battery too low to complete screening Step 8 Observe screening complete At the completion of the screen the ALGO 3i screener will save a...

Page 11: ...reening in Progress window i If impedances are between 8 kOhms and 11 kOhms the Impedances Marginal message will appear showing actual impedances and asking you to confirm your intention to begin scre...

Page 12: ...his area Ensure that all infant data is accurate before clicking the OK key e If you do not wish to enter infant data for the screening record proceed with screening At the completion of the screening...

Reviews: