background image

5.  Position the objective turret until the 4x (smallest) objective lens “clicks” into position in the optical path.  Note that 

each time you change from one objective lens to another you should turn the turret until you hear the “click”, which 
indicates that the lens is properly indexed in the optical path. 

 

6.  Swing moveable finger on slide holder outward.  Place specimen slide (cover slip up) on top of stage surface against 

fixed side of slide holder.  Slowly release moveable finger until it makes contact with specimen slide. 

 

7.  Turn rheostat (dimmer) control in direction to reduce illumination brightness to about mid-point. Normally, brightness 

will need to be reduced as magnification is reduced, and increased as magnification is increased.  Experiment with 
this adjustment until optimum resolution and contrast of specimen is obtained. 

 

8.  Locate lever beneath stage that controls the iris diaphragm.  Move lever until iris diaphragm is about half open.  

Normally, aperture of iris should become smaller for lower magnifications and larger for higher magnifications.  
Experiment with this setting until optimum resolution and contrast of specimen is obtained.   

 

9.  While looking through the eyepiece, rotate coarse focusing knobs until specimen comes into focus. If image does not 

appear in field of view, move specimen slide slightly until image appears in field of view. 

 

10.  Adjust fine focus controls until specimen is in sharp focus. 

 

11.  Changing magnification. 

 

a.  Magnification is changed by rotating objective turret until a different objective lens is moved into optical path.  

Always turn turret until you hear the “click”, indicating that lens is properly indexed.  Otherwise, you will not be 
able to see anything when looking through the microscope. 

 

Objective Specifications 

 

Objective 

N.A. 

Color Code 

Ring 

Field of 

View 

Working 

Distance 

Magnification with 

WF10X eyepiece 

Din 4X 

0.10 

Red 

4.5mm 

18.5mm 

40X 

Din 10X 

0.25 

Yellow 

1.8mm 

6.4mm 

100X 

Din 40X retractable 

0.65 

Blue 

0.45mm 

0.5mm 

400X 

Din 100x retractable 

1.25 

White 

0.18mm 

0.06mm 

1000X 

 

b.  Also note that each objective has a color ring, which permits you to instruct changes in magnification by referring 

to an easily observed color rather than to a number. 

 

c.  The microscope has been parfocalled at the factory, which allows easy change from one magnification to another, 

requiring little or no adjustment of the fine focus knobs. 

 

d.  As magnification is increased, the field of view (area of specimen seen through the microscope) will decrease.  

That is why it is easier to find the specific area of interest on the specimen by starting with the lowest 4x objective 
lens, before increasing magnification with the higher power objective lens. 

 

e.  NOTE:  Care must be taken when rotating the 40x and 100x objective lenses into place.  These lenses have a 

spring retractable mechanism, which retracts slightly into its housing if the front of the lens, strikes the specimen 
slide.  With fine focus adjustment at mid-range, the rack stop has been adjusted at the factory to assure the 40x 
and 100x lens will clear the thickness of a normal specimen slide and cover slip.  However, if the rack stop has 
been improperly adjusted, or if using a thicker than normal slide or cover slip, moving the 40x and 100x lenses too 
quickly or carelessly could cause damage to the front lens element or to the slide. 

 

f. 

In order to obtain maximum resolution of the 100x oil immersion lens, it is necessary to apply immersion oil 
between the cover glass of slide and front lens of the objective. Use a very small amount of immersion oil.  All air 
bubbles must be removed from between lens and slide by gently rotating nosepiece back and forth. 

 

When you are finished viewing, all parts that come in contact with oil must be cleaned.  Failure to do so could 
permanently damage the 100x immersion objective lens. 

 

Do not let the front lens element of the 40x objective lens come into contact with a wet slide surface, as prolonged 
contact with any moisture could damage the lens.  If lens is exposed to moisture, promptly wipe with soft tissue to 
remove moisture. 

 

 

 

Summary of Contents for DCX-213-RLED

Page 1: ...DCX 213 RLED MODEL DCX 214 RLED COMPOUND BIOLOGICAL MICROSCOPE WITH WIFI CAMERA Manual for microscope and WiFi camera operation only Microscope serial number This number is on a label located under t...

Page 2: ...Specimen holder mechanical stage Viewing head of microscope Widefield 10x 18 eyepiece Head locking screw Arm of microscope Coaxial focus knob Tension adjustment collar behind focus knob on left side...

Page 3: ...ng in the cameras IP address in your web browser These instructions are covered in WiFi Camera Operation section of this manual 4 If you would like to connect the WiFi camera to a wireless enabled lap...

Page 4: ...eplacing and each charge will provide up to 50 hours of microscope operation The LED component bulb will last for up to 50 000 hours before replacement is required WARNING DO NOT USE regular AA alkali...

Page 5: ...in 40X retractable 0 65 Blue 0 45mm 0 5mm 400X Din 100x retractable 1 25 White 0 18mm 0 06mm 1000X b Also note that each objective has a color ring which permits you to instruct changes in magnificati...

Page 6: ...will open the Live Imaging Module Locate the Video Device box You will notice that by default the Moticam X is selected Click on the Open button and the software will enable the camera If you are usi...

Page 7: ...nd charger automatically switches to trickle charge The charger can be left plugged in but for safety reasons it is a good idea to disconnect the charger from the AC wall outlet and the output cord fr...

Page 8: ...ld of view Eyepiece or condenser lens dirty Clean lens Specimen slide dirty Clean slide No WiFi signal No power to the camera Make sure blue LED is solid Make sure camera power adapter is plugged in M...

Reviews: