Appendix A Specifications for VXI-MIO-64E-1
VXI-MIO Series User Manual
A-2
National Instruments Corporation
Input coupling....................................DC
Max working voltage
( common mode).................Each input should remain within
±
11 V of ground
Overvoltage protection.......................
±
25 V powered on,
±
15 V
powered off
Inputs protected .......................... ACH<0..63>, AISENSE,
AISENSE2
FIFO buffer size.................................8,192 S
Data transfers .....................................DMA, interrupts, programmed
I/O
Configuration memory size ................512 words
Input signal ranges ................
Module Gain
(Software
Selectable)
Module Range
(Software Selectable)
Bipolar
Unipolar
0.5
±
10 V
—
1
±
5 V
0 to 10 V
2
±
2.5 V
0 to 5 V
5
±
1 V
0 to 2 V
10
±
500 mV
0 to 1 V
20
±
250 mV
0 to 500 mV
50
±
100 mV
0 to 200 mV
100
±
50 mV
0 to 100 mV