DMM Isolation Voltages

Hazardous Voltage

 

This icon denotes a warning advising you to take precautions to avoid electrical shock.

Channel-to-earth ground

Continuous

300 V, Measurement Category II

Withstand

3,000 V RMS, verified by a 5 s dielectric withstand test

Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level

electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Caution

 

Do not connect the VirtualBench hardware to signals or use for measurements within Measurement Categories III or IV.

DC Power Supply Isolation Voltages

+25 V and -25 V-to-earth ground, continuous

60 V DC, Measurement Category I

Note

 

Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct

connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Power Requirements

Caution

 

The protection provided by the VirtualBench hardware can be impaired if it is used in a manner not described in the 

NI VirtualBench

VB-8054 Safety, Environmental, and Regulatory Information

.

Voltage input range

100 V AC to 240 V AC, 50/60 Hz

Power consumption

150 W maximum

Power input connector

IEC C13 power connector

Power disconnect

The AC power cable provides main power disconnect. Do not position the

equipment so that it is difficult to disconnect the power cable. Depressing the

front panel power button does not inhibit the internal power supply.

Preparing the Environment

Ensure that the environment you are using the VB-8054 in meets the following specifications.

Operating temperature

0 °C to 40 °C

Cooling

Forced air circulation (negative pressurization) through a fan. Fan speed

automatically adjusts according to operating conditions. Intake locations are on

the sides of device. Exhaust location is on the rear of device. Ensure that the

intake and exhaust locations are not obstructed.

Pollution Degree

2

Maximum altitude

2,000 m

Indoor use only.

Note

 

Refer to the 

NI VirtualBench VB-8054 Specifications

 for complete specifications.

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications.

These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational

electromagnetic environment.

This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a

peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and

prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.

Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your

local regulatory rules.

Caution

 

To ensure the specified EMC performance, operate this product only with shielded cables and accessories connected to the Ethernet, USB,

FGEN, TRIG, and oscilloscope CH 1 though 4 ports.

2

  | 

ni.com

  | 

VB-8054 Safety, Environmental, and Regulatory Information