Environment
Maximum altitude
4,600 m operational
Pollution Degree
2
Indoor use only.
Operating Environment
Ambient temperature range
0 °C to 50 °C (Tested in accordance with
IEC 60068-2-1 and IEC 60068-2-2.)
Relative humidity range
14
(noncondensing)
At 50 °C
45%
At 40 °C
75%
At 30 °C
95%
Storage Environment
Ambient temperature range
-40 °C to +71 °C (Tested in accordance with
MIL-PRF-28800F (Class 3).)
Relative humidity range
5% to 95%, noncondensing (Tested in
accordance with MIL-PRF-28800F (Class 3).)
Shock and Vibration
Operating shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating
5 Hz to 500 Hz, 0.3 g
rms
Nonoperating
5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
14
Tested in accordance with IEC 60068-2-56.
USB-5683 Specifications
|
© National Instruments
|
9