background image

Exclusion periods

11

Start exclusion

0 ms to 10 ms

End exclusion

0 ms to 10 ms

Trigger

Internal Trigger

Source

12

Bus or Continuous (Auto, Single or Multiple)

Range

-35 dBm to +20 dBm

Level accuracy

±0.5 dB, typical

Slope

Positive or negative

Delay range

-5 ms to 10 s

Delay resolution

10 μs

Trigger In

Impedance

4 kΩ, nominal

Type

TTL/CMOS

Slope

Positive

Delay range

-5 ms to 10 s

Delay resolution

10 μs

Voltage high threshold

2.3 V, typical

Voltage low threshold

1.2 V, typical

Hysteresis

0.2 V, typical

Maximum voltage

±5.5 V

Minimum pulse width

100 ns

Latency

13

10.6 μs, maximum

Repetition period

14.2 μs, minimum

Trigger Out

Trigger Out capability is not currently implemented.

11

The start exclusion time plus the end exclusion time must be less than the slot width.

12

Software trigger not available in the NI-568x SFP. This feature is only available when using the
NI-568x instrument driver. Internal and external triggers are not available when using Continuous
acquisition mode.

13

Latency is defined as the time delay between the sensor receiving the trigger and the measurement
initiation.

USB-5683 Specifications

  | 

© National Instruments

  | 

7

Summary of Contents for USB-5683

Page 1: ...8 DC Power Requirements 5 V from Host USB 8 Calibration 8 Physical Characteristics 8 Environment 9 Operating Environment 9 Storage Environment 9 Shock and Vibration 9 Compliance and Certifications 10...

Page 2: ...ns are Typical unless otherwise noted Conditions Minimum or maximum specifications are warranted under the following conditions unless otherwise noted 1 hour warm up time at ambient temperature Calibr...

Page 3: ...surement Uncertainty dB 0 05 GHz Initial Power Measured dBm Final Power Measured dBm 0 C to 50 C 20 C to 30 C 4 to 20 16 to 4 60 to 16 4 to 20 16 to 4 60 to 16 60 to 16 0 14 0 13 0 03 0 08 0 09 0 03 1...

Page 4: ...123 pW 332 pW 344 pW 16 to 4 10 1 nW 38 7 nW 42 9 nW 4 to 20 0 856 W 1 07 W 0 996 W 3 Noise is specified as two sigma at three sigma confidence after zero operation when measured with 10 2 seconds of...

Page 5: ...ta points Continuous Mode Duty cycle correction 0 001 to 100 Aperture time 0 01 ms to 1 s 6 Measurement error with reference to a CW signal of equal power and frequency at 20 C to 30 C 7 Maximum numbe...

Page 6: ...e Maximum number of slots 128 Slot width 0 01 ms to 100 ms Maximum capture time 1 000 ms Slot Width Number of Slots Resolution 0 01 maximum 9 Times are typical based on benchmark results taken using a...

Page 7: ...al Hysteresis 0 2 V typical Maximum voltage 5 5 V Minimum pulse width 100 ns Latency13 10 6 s maximum Repetition period 14 2 s minimum Trigger Out Trigger Out capability is not currently implemented 1...

Page 8: ...mA Caution You can impair the protection provided by the USB 5683 if you use it in a manner not described in this document Calibration Interval 1 year calibration interval starts with the date the pro...

Page 9: ...MIL PRF 28800F Class 3 Relative humidity range 5 to 95 noncondensing Tested in accordance with MIL PRF 28800F Class 3 Shock and Vibration Operating shock 30 g peak half sine 11 ms pulse Tested in acco...

Page 10: ...CISPR 22 Class A emissions FCC 47 CFR Part 15B Class A emissions ICES 001 Class A emissions Note In the United States per FCC 47 CFR Class A equipment is intended for use in commercial light industri...

Page 11: ...s For additional environmental information refer to the Minimize Our Environmental Impact web page at ni com environment This page contains the environmental regulations and directives with which NI c...

Page 12: ...EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and h...

Reviews: