TC-2095 Terminal Block Calibration Procedure
2
ni.com
Equipment and Other Test Requirements
This section describes the equipment, software, documentation, and test
conditions required for verifying the performance of the TC-2095.
Test Equipment
Verification requires a high-precision voltage source with at least 50 ppm
accuracy, a multiranging 5 1/2 digit digital multimeter (DMM) with
15 ppm accuracy, and a thermometer that is accurate to within 0.1
°
C.
NI recommends you use the following instruments for verifying the
performance of the TC-2095:
•
Calibrator—Fluke 5700A
•
DMM—NI 4060 or Agilent 34401A
Note
You also need three test connector adapters that fit the 96-pin connector pins.
Use a Pomona 3560 banana jack to test adapter connector (0.040 in. diameter pins) or
an equivalent connector.
If these instruments are not available, use the accuracy requirements listed
above to select a substitute calibration standard.
Software and Documentation
You can find all the necessary information to verify the performance
of the TC-2095 in this verification procedure. No other software or
documentation is required. If you would like more information on the
TC-2095, refer to the
BNC/TC-205 Rack-Mount Adapter Installation
Guide
, which you can download from the NI Web site at
ni.com/manuals
.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during verification:
•
Keep connections to the TC-2095 terminal block short. Long cables
and wires act as antennae, picking up extra noise that can affect
measurements.
•
Use shielded copper wire for all cable connections to the terminal
block. Use twisted-pair wire to eliminate noise and thermal offsets.
•
Keep relative humidity below 80%.
•
Maintain a temperature between 15 and 35
°
C.