National Instruments Corporation
7
TBX-1316 High-Voltage Attenuator Terminal Block
5.
Connect chassis ground to the chassis-ground terminal strip.
6.
Connect the signal wires to the signal wire screw terminals by inserting
the stripped end of the wire fully into the terminal.
Caution
No bare wire should extend past the screw terminal. Exposed wire increases the
risk of short circuits and failures.
7.
Tighten the terminal screws to a torque of 1.2 to 1.5 N · m
(10.62 to 13.28 in. · lb).
8.
Tighten the signal strain-relief collar.
9.
Repeat steps 2 through 8 for each signal wire.
10. Close and lock the enclosure door.
11. Connect the TBX-1316 to the SCXI module front connector as
explained in the
Connecting the Module
section.