Appendix D
Common Questions
D-2
ni.com
Can I use the unused analog input channels of the E/M Series DAQ
device if I am directly cabled to the SCXI-1102/B/C, for example with
the SCXI-1180 feedthrough?
Yes. E/M Series DAQ device channels 1 through 7 are available when
directly connected to the SCXI-1102/B/C.
Which digital lines are unavailable on the E/M Series DAQ device if I
am cabled to an SCXI-1102/B/C module?
Table D-1 shows the digital lines that are used by the SCXI-1102/B/C for
communication and scanning. These lines are unavailable for
general-purpose digital I/O if the SCXI-1102/B/C is connected to the
E/M Series DAQ device.
Table D-1.
Digital Signals on the SCXI-1102/B/C
E/M Series
DAQ Device
Signal Name
NI-DAQmx
SCXI Signal
Name
Traditional
NI-DAQ
(Legacy) SCXI
Signal Name
50-Pin
Connector
68-Pin
Connector
Direction
1
DIO0
P0.0
SERDATIN
25
52
Output
DIO4
P0.4
SERDATOUT
26
19
Input
DIO1
P0.1
DAQD*/A
27
17
Output
DIO2
P0.2
SLOT0SEL*
29
49
Output
SCANCLK
AI Hold Comp,
AI Hold
SCANCLK
36
46
Output
PFI7/
STARTSCAN
PFI 7/
AI SAMP CLK,
AI SAMP
HOLDTRIG
46
38
Input
EXTSROBE*
EXTSROBE*
SERCLK
37
45
Input
1
With respect to the E/M Series DAQ device.