SCXI-1324 Installation Guide
4
www.ni.com
4.
Prepare your signal wire by stripping the insulation no more than
7 mm.
5.
Connect the wires to the screw terminals by inserting the stripped end
of the wire fully into the terminal. No bare wire should extend past the
screw terminal. Exposed wire increases the risk of short circuiting
adjacent signals causing a circuit failure.
6.
Tighten the screws to a torque of 5–7 in.-lb.
7.
Connect safety earth ground to the safety-ground solder lug. Refer to
the
section for connection information.
8.
Reinstall the strain-relief bar and tighten the strain-relief screws.
9.
Reinstall the top cover and tighten the top cover screws.
10. Connect the terminal block to the module front connector as explained
in the
Connecting the Terminal Block to the SCXI Module
Figure 1. SCXI-1324 Parts Locator Diagram
1
Strain-Relief Bar
2
Strain-Relief Screws
3
Safety-Ground
Solder Lug
4
Mating Connector
5
Thumbscrew
6
Top Cover Screws
7
Top Cover
7
6
5
2
3
4
1
Back View
Front View
Summary of Contents for SCXI-1324