Chapter 2
Using the SCXI-1129
©
National Instruments Corporation
2-17
The matrix expansion methods available for the different terminal blocks
are shown in Table 2-2.
The preferred expansion method is to connect columns using the matrix
expansion plug. You can use this plug to daisy-chain module columns
across an entire chassis, as shown in Figure 2-12.
Caution
Only daisy-chain terminal blocks of the same type together. Daisy-chaining
different types of terminal blocks together can cause damage to the module or to the
equipment connected to the module. For example, you can daisy-chain two or more
SCXI-1333 terminal blocks; however, do
not
daisy-chain the SCXI-1333 together with any
other type of terminal block.
Table 2-2.
Terminal Block Matrix Expansion
Terminal Block
Expansion Type
SCXI-1333 (4
×
16)
Column connection
SCXI-1337 (8
×
16)
None
SCXI-1339 (4
×
32)
None
SCXI-1334 (4
×
64)
Column and row connection
SCXI-1335 (8
×
32)
Column and row connection
SCXI-1336 (16
×
16)
Column connection