SCXI-1120/D Calibration Procedure
Adjusting Offset Null Values of the Module
Complete the following steps to adjust the offset null value:
1.
Set the channel gain on all channels to a gain of 1. Set the filter value
to 4 Hz for the SCXI-1120 module or 4.5 kHz for the SCXI-1120D
module. Refer to the
section in this document
for information on how to set the channel gain.
2.
Connect the calibrator to the analog input channel you want to adjust,
starting with channel 0. Refer to Table 5 to determine the pins on the
96-pin front connector that correspond to the positive and negative
inputs of the specified channel. For example, the positive input for
channel 0 is pin A32, which is labeled
CH0+
. The negative input for
channel 0 is pin C32, which is labeled
CH0–
.
3.
Connect the DMM to the output of the same channel to which the
calibrator was connected in step 2. Refer to Figure 4 to determine the
pins on the 50-pin rear connector that correspond to the positive and
negative outputs for the specified channel. For example, the positive
output for channel 0 is pin 3, which is labeled
MCH0+
. The negative
output for channel 0 is pin 4, which is labeled
MCH0–
.
4.
Set the calibrator to output 0.0 V.
5.
Adjust the output potentiometer of the channel until the DMM reading
is 0 ±3.0 mV. Refer to Figure 3 for the potentiometer location and
Table 6 for the potentiometer reference designator.
6.
Set the channel gain on all channels to 1000.0.
Table 6.
Calibration Potentiometers Reference Designators
Input Channel Number
Input Null
Output Null
0
R08
R24
1
R10
R25
2
R12
R26
3
R14
R27
4
R16
R28
5
R18
R29
6
R20
R30
7
R21
R31