National Instruments SCXI-1104 Manual Download Page 6

SCXI-1104 Calibration Procedure

6

ni.com

DAQChannel

0

DAQgain

—0 for 16-bit devices, –1 for 12-bit devices

numPoints

—100

binArray

—The array returned from the

DAQ_Op

function

The result is a set of scaled voltages that you read from the SCXI
module.

12. Average the results returned from the

SCXI_Scale

function. Compare

the averaged result to the upper and lower limits in Table 1. If the result
falls within these limits, your module has passed this portion of the
test. If it does not, use the steps in the next section,

Adjusting Gain and

Offset Errors and Calibration Constants

, to adjust your module.

13. Repeat steps 7 through 12 for the remaining channels.

You have now completed verifying the operation of your SCXI-1104.

Note

If all channels fall within the upper and lower limits, your module does not need to

be adjusted.

Adjusting Gain and Offset Errors and Calibration Constants

This section contains step-by-step instructions on adjusting gain and offset
errors and calibration constants.

Adjusting Gain and Offset Errors

Complete the following steps to adjust gain and offset errors in your
SCXI-1104:

1.

Read the

Test Conditions

section.

2.

Connect the E Series device to the SCXI-1104.

3.

Call

Calibrate_E_Series

to minimize the uncertainty associated

with the E series device. Set the following parameters:

deviceNumber

—The value obtained from Measurement &

Automation Explorer

calOp

ND_SELF_CALIBRATE

setOfCalConstants

ND_USER_EEPROM_AREA

calRefVolts

—0.0

4.

Call

AI_Configure

to put the E Series device in bipolar mode. Set the

following parameters:

deviceNumber

—The value obtained from Measurement &

Automation Explorer

chan

—0

Summary of Contents for SCXI-1104

Page 1: ...rcuitry that corrects these errors National Instruments calibrates every SCXI 1104 module at the factory During the factory calibration process the calibration constants are stored in the non volatile...

Page 2: ...need a connector block such as the National Instruments SCXI 1300 a shielded 68 pin connector cable a 50 pin ribbon cable a 50 pin breakout box and an SCXI adapter These components give easy access to...

Page 3: ...w these guidelines to optimize the connections and the environment during calibration Keep connections to the SCXI module as short as possible Long cables and wires can act as antennae which pick up e...

Page 4: ...ete the following steps to verify the operation of your SCXI 1104 1 Read the Test Conditions section earlier in this document 2 Refer to Table 1 at the end of this document to see the test specificati...

Page 5: ...tor to the CH0 and CH0 inputs 8 Call SCXI_Single_Chan_Setup to configure the module for single channel measurements Set the following parameters SCXIchassisID The value obtained from Measurement Autom...

Page 6: ...XI 1104 Note If all channels fall within the upper and lower limits your module does not need to be adjusted Adjusting Gain and Offset Errors and Calibration Constants This section contains step by st...

Page 7: ...positive and negative inputs of the specified channel For example the positive input for channel 0 is pin C32 which is labeled CH0 The negative input for channel 0 is pin B32 which is labeled CH0 If...

Page 8: ...and binary2 with volt2 for later use 15 Repeat step 4 through 14 for the remaining channels you want to adjust Adjusting Calibration Constants Complete the following steps to adjust the calibration co...

Page 9: ...on the SCXI 1104 calConst1 and calConst2 These are return values 2 Repeat step 1 changing calibrationArea to 1 3 Repeat step 1 changing calibrationArea to 3 4 Repeat steps 1 through 3 for the next cha...

Page 10: ...NC NC CGND 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 A CH31 CH30 CH29 CH28 CH27 CH26 CH25 CH24 CH23 CH22 CH21 CH20 CH19 CH18 CH17 CH16 CH15 CH14 CH13 CH12...

Page 11: ...er Limit and Lower Limit values Table 1 SCXI 1104 Specifications Ranges Gain Test Point V Upper Limit V Lower Limit V 0 100 41 8000 41 8219 41 7781 0 0000 0 0010 0 0010 41 8000 41 7781 41 8219 NC NC N...

Reviews: