background image

USER GUIDE

SCB-19

19-Pin Shielded Desktop Connector Block

This document describes how to connect and use the SCB-19 with the AUX I/O front panel 
connector on your National Instruments device. Figure 1 shows the parts of the 
SCB-19 connector block.

Figure 1.  

SCB-19 Connector Block

1

SCB-19

2

Cover Screws

1

2

2

Summary of Contents for SCB-19

Page 1: ...lock This document describes how to connect and use the SCB 19 with the AUX I O front panel connector on your National Instruments device Figure 1 shows the parts of the SCB 19 connector block Figure 1 SCB 19 Connector Block 1 SCB 19 2 Cover Screws 1 2 2 ...

Page 2: ...pter module or other NI device with compatible 19 pin AUX I O front panel connector SHH19 H19 AUX cable assembly NI part number 152629 0x Phillips 1 screwdriver 0 125 in flathead screwdriver Shielded multiconductor cable with 14 30 AWG wire Wire cutters Wire insulation stripper Note These supplies are not included in the SCB 19 kit You must purchase these items separately ...

Page 3: ...erminals and securely tightening the screws with the flathead screwdriver to a torque of 0 5 N m to 0 6 N m 4 43 lb in to 5 31 lb in Caution To ensure the EMC performance specified for the connected device any wires connected to screw terminals that exit the enclosure must be shielded NI recommends using a multiconductor cable with an overall shield Terminate the cable shield to one of the PCB mou...

Page 4: ...any damage resulting from using voltages outside of the recommended range 5 Connect the SCB 19 to the AUX I O front panel connector on your NI device using an SHH19 H19 AUX cable NI part number 152629 0x and tighten the target screw as shown in Figure 3 Figure 3 SCB 19 Cable Connection Caution Although the AUX I O front panel connector accepts a standard HDMI cable a standard HDMI cable causes sig...

Page 5: ...781740 01 The SCB 19 also has two screw holes for a generic wall mount These wall mount holes are designed for use with a 4 or 6 panhead screw with a minimum length of 0 625 in 15 88 mm Figure 4 shows the screw holes locations on the bottom of the SCB 19 Figure 4 SCB 19 Mounting Screw Holes 1 DIN Rail Mounting Screw Holes 2 Wall Mounting Screw Holes 40 Prod of U S A M N SCB 19 P N 151899A 01L S N ...

Page 6: ... accordance with IEC 60068 2 1 and IEC 60068 2 2 Relative humidity range 5 to 95 noncondensing in accordance with IEC 60068 2 56 Note Clean the device with a soft non metallic brush Make sure that the device is completely dry and free from contaminants before returning it to service Shock and Vibration Operational shock 30 g peak half sine 11 ms pulse in accordance with IEC 60068 2 27 Test profile...

Page 7: ...g and manufacturing products in an environmentally responsible manner NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers For additional environmental information refer to the NI and the Environment Web page at ni com environment This page contains the environmental regulations and directives with which NI complies as w...

Page 8: ...ort compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Support The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance f...

Reviews: