RM-4339 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification procedures. If the
recommended equipment is not available, select a substitute using the requirements listed in
Table 1.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Requirements
Calibrator
Fluke 5520A
If this instrument is unavailable, use a
calibrator that can provide resistance
values in the range of 120
Ω
to 1 k
Ω
with
0.01
Ω
resolution, and an accuracy of
90 ppm or better, and can sink 15 mA on
the 120
Ω
setting, 8 mA on the 350
Ω
setting, and 3 mA on the 1 k
Ω
setting.
Ohm Meter
NI PXI-4070
If this instrument is unavailable, use a
6 ½ digit DMM with 2-wire resistance
measurement capability and resistance
accuracy of 80 ppm or better of reading
+6 ppm of range for the 100 k
Ω
range.
8-Channel
Universal-Bridge Input
Module
NI PXIe-4339
—
Terminal Block
Connection Accessory
TB-4339/B/C
—
PXI Express Chassis
NI PXIe-1062Q
—
Accessory Cable
SH96-96-2
The SH96-96-2 is available in 1 meter,
3 meter and 5 meter lengths. Any of these
cable lengths can be used.
RJ50 to Pigtail Cable (8)
NI P/N: 195950-02
An RJ50 cable with a cable wire
resistance < 0.2
Ω
with better than 10%
resistance mismatch between individual
wires.