6
|
ni.com
|
RM-4339 Calibration Procedure
3.
Create a task to enable SCA shunt calibration.
a.
Create and configure an AI Bridge (V/V) measurement according to the values found
in Table 2.
b.
Enable the SCA shunt calibration circuit using the following parameters:
Property Node: DAQmx Channel
AI.Bridge.ShuntCal.Select: A
AI.Bridge.ShuntCalASrc: Built-In
AI.Bridge.ShuntCal.ShuntCalAResistance: 100 k
AI.Bridge.ShuntCal.Enable: True
Note
These properties are located at
Analog Input»General Properties»Signal
Conditioning»Bridge»Shunt Cal
.
c.
Commit the task by calling DAQmx Control Task with the following parameter:
Action: Commit
4.
Take a resistance measurement from the DMM and record the result.
5.
Compare the result from step 4
to the Upper Limit and Lower Limit values in Table 3. If
the result is between these values, the module passes the test.
6.
Clear the task by calling DAQmx Clear Task.
Table 2.
AI Bridge Configuration
Configuration
Value
Channel Name
Dev1/aix
,
where
x
refers to the channel number.
Measurement Type
AI Bridge (V/V)
Voltage Excitation Source
Internal
Voltage Excitation Value
3.3
Maximum Value
0.01
Minimum Value
-0.01
Units
V/V
Bridge Configuration
Full
Sample Mode
Finite Samples
Sample Clock Rate
100
Samples per Channel
100