2. Launch MAX and perform an intermediate self-test.
1. Select the item in the configuration tree and click
Self-Test
in the MAX
toolbar.
2. Select the
Intermediate
radio button and then click
Start
.
The green signal illuminates and shows
Pass
when the self-test completes
successfully.
If the self-test fails, review the
Detailed Result
pane in the 5842 Self-Test
window to identify the specific item(s) causing the failure, and inspect any
cables and connections for the item(s). Address any issues you identify and
perform an intermediate self-test again. If the self-test still fails, continue this
procedure.
3. Power
off
the chassis.
4. Reinstall the failed instrument in a
different
set of slots.
5. Power on the chassis.
6. Perform an intermediate self-test again.
What Should I Do if the Instrument Does Not
Initialize?
Failure to initialize may indicate a problem with module interconnections or with
MAX.
If the niRFSA Initialize VI or niRFSG Initialize VI returns an error and the PXIe-5842
fails to initialize, complete the following steps:
1. Power
off
the system.
2. Inspect all PXIe-5842 front panel cables and, if there are no apparent issues,
reconnect cables securely, according to the interconnection procedure for
your instrument.
3. Power on your system and run the MAX configuration and intermediate self-
test procedures.